Back to results

Virginia Commonwealth University

Integration of VMC++ into a Commercial Treatment Planning System

Abstract

dc:description.abstract

Recently, there has been interest to integrate VMC++ into the commercial treatment planning system at VCU as another Monte Carlo code option, since it has been shown to increase efficiency dramatically without introducing a significant amount of systematic error. Also, independent validation of VMC++ for photon beams is of interest since this has not been performed previously in literature. This study included several tests required to integrate VMC++. Output factor normalization was performed and found to agree with experiment to within 1% for all field sizes except 1x1 cm2. Geometric validation was successful. Dosimetric validation was performed with respect to DOSXYZnrc on a water phantom, resulting in agreement within statistical uncertainty except for slight differences at the surface of the phantom. Dosimetric comparison was made for a head-and-neck patient case, showing that 5% of the voxels did not agree within 2.8% of maximum dose. The ability of VMC++ to compute dose-to-water was compared to an in-house algorithm and found to agree within statistical uncertainty.

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
Thesis
Discipline thesis:degree_discipline
Physics
Year dc:date.available
2005

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Gardner, Joseph Kingsley
Contributors dc:contributor
  • Dr. Jeffrey V. Siebers

Subjects

dc:subject × 12

Rights

dc:rights
Statement dc:rights
  • © The Author

Identifiers

dc:identifier.*
OAI identifier oai:identifier
oai:scholarscompass.vcu.edu:etd-1989

Chain of custody

source
Harvested from
Virginia Commonwealth University
Base URL
scholarscompass.vcu.edu/do/oai/
Last updated
2026-07-24
Source record
OAI-PMH GetRecord
citation

Gardner, Joseph Kingsley. Integration of VMC++ into a Commercial Treatment Planning System. Thesis thesis, 2005. https://doi.org/10.25772/KJCB-B414