Back to results

University of Illinois at Urbana-Champaign

Joint-optimization of the x-ray source voltage, beam filtration and detection angle for a full-ring benchtop x-ray fluorescence computed tomography (XFCT) image system

Abstract

dc:description

In this thesis, we present a combined theoretical and Monte Carlo study to optimize the design of a full-ring desktop X-Ray Fluorescence Computed Tomography (XFCT) system based on ultrahigh energy resolution CdTe detectors and a pencil-beam-irradiation/multiple-slit camera (PB/MS) geometry. In this study, we have specifically focused on a joint-optimization of X-ray source voltage, beam filtering and signal-collection angle for improving the signal-to-noise ratio attainable with the imaging setup. Given a large number of possible combinations of these design parameters, we have developed an analytical approach that could be used to rapidly evaluate the intrinsic signal-to-noise ratio (SNR) that measures the fluorescence signal, Compton scattering noise, and Rayleigh scattering noise generated under a simplified X-ray irradiation and data collection geometry. We have used this method to search through a large number (a total of 4800 in this study) of different combinations of system parameters, such as X-ray tube voltage, filter material, thickness, angle of detection, etc., to identify the potential optimum choice of system parameters. The study shows the following two conclusions, (1) If the dose is the primary concern, the value of the SNR improved with an increased filter thickness and a large back-ward signal collection angle, (2) Given the X-ray tube maximum power, the maximum tube voltage with a signal collection angle of 80-100 degrees can reach the optimum SNR. These results were further validated through Monte Carlo studies with more realistic imaging geometries.

Degree

thesis:*
Name thesis:degree_name
M.S.
Level thesis:degree_level
Thesis
Discipline thesis:degree_discipline
Nuclear, Plasma, Radiolgc Engr
Grantor
University of Illinois at Urbana-Champaign
Year dc:date
2022

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Nie, Xingchen
Contributors dc:contributor
  • Meng, Lingjian
  • Di Fulvio, Angela

Subjects

dc:subject × 4

Rights

dc:rights
Statement dc:rights
  • Copyright 2022 Xingchen Nie
Language dc:language
en, eng

Identifiers

dc:identifier.*
Handle dc:identifier
https://hdl.handle.net/2142/116165

Chain of custody

source
Harvested from
University of Illinois - Urbana-Champaign
Base URL
www.ideals.illinois.edu/oai-pmh
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
citation

Nie, Xingchen. Joint-optimization of the x-ray source voltage, beam filtration and detection angle for a full-ring benchtop x-ray fluorescence computed tomography (XFCT) image system. Thesis thesis, University of Illinois at Urbana-Champaign, 2022. https://hdl.handle.net/2142/116165