Abstract
dc:descriptionAccessories have been designed and constructed which directly transform a conventional single-beam spectrophotometer into a versatile stopped-flow kinetic analyzer. A relatively inexpensive microcomputer system serves as a dedicated controller/data processor for the instrument. The design employed features preservation of the optical configuration and geometry of the original spectrophotometer. Thus, abilities related to the performance of precise, absolute absorbance measurements have been retained. Additionally, kinetic response curves can be reproducibly generated, primarily due to efficiencies on solution transport and mixing components built into the "stopped-flow module". The hybrid nature of the instrument, allowing reliable collection and processing of responses from both kinetic and equilibrium phases of reactions, is seen as a powerful attribute.
Degree
thesis:*- Name thesis:degree_name
- Ph.D.
- Level thesis:degree_level
- Dissertation
- Discipline thesis:degree_discipline
- Chemistry
- Grantor
- University of Illinois at Urbana-Champaign
- Year dc:date
- 2014
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Seiler, Brian Douglas
Subjects
dc:subject × 1Identifiers
dc:identifier.*- Identifier
- (UMI)AAI8324641
- OAI identifier oai:identifier
- oai:www.ideals.illinois.edu:2142/70227