Elastic-anisotropic properties of the porcine superflexor tendon measured by instrumented indentation

Abstract

dc:description

Measurements of soft tissue anisotropy using instrumented indentation are of interest because of the potential for indentation-based clinical diagnostics. Shortage of previous research relating to instrumented indentation of soft tissue in the kPa range, such as tendon, motivated a protocol development phase in our study. Once indents in tendon were repeatedly successful and the protocol validated, elastic-anisotropy in tendon was studied on the premise that collagen fibers in tendon are highly-ordered and unidirectional. Samples of porcine superflexor tendon were mounted for indentation such that the collagen fibers were oriented one of three possible ways relative to the indentation probe: parallel to the probe (longitudinal indentation), perpendicular to the probe (lateral indentation), or 45° to the probe (intermediate indentation). Non-significant differences in the reduced elastic modulus of the three experimental groups suggest that porcine superflexor tendon anisotropy is not detected by instrumented indentation. Contextualizing these results within the concept of indentation strain provides support for the possibility that the critical strain required for the onset of anisotropy in tendon is perhaps beyond the deformation range on the instrumented indentation system. With respect to protocol development for instrumented indentation of soft tissues, new insight is provided into the importance of clearing the indentation probe between successive indents of biological debris to ensure data integrity.

Degree

thesis:*
Name thesis:degree_name
M.S.
Level thesis:degree_level
Thesis
Discipline thesis:degree_discipline
Mechanical Engineering
Grantor
University of Illinois at Urbana-Champaign
Year dc:date
2014

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Sangha, Harpreet
Contributors dc:contributor
  • Wagoner Johnson, Amy J.

Subjects

dc:subject × 5

Rights

dc:rights
Statement dc:rights
  • Copyright 2014 Harpreet Sangha
Language dc:language
en

Identifiers

dc:identifier.*
Handle dc:identifier
http://hdl.handle.net/2142/49786
OAI identifier oai:identifier
oai:www.ideals.illinois.edu:2142/49786

Chain of custody

source
Harvested from
University of Illinois - Urbana-Champaign
Base URL
www.ideals.illinois.edu/oai-pmh
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
citation

Sangha, Harpreet. Elastic-anisotropic properties of the porcine superflexor tendon measured by instrumented indentation. Thesis thesis, University of Illinois at Urbana-Champaign, 2014. http://hdl.handle.net/2142/49786