Back to results

Massachusetts Institute of Technology

Soft pneumatic artificial muscles with low threshold pressures for a cardiac compression device

Abstract

dc:description.abstract

In this paper, I present the design, fabrication and characterization of fully soft pneumatic artificial muscles (PAMs) with low threshold pressures that are intended for direct cardiac compression (DCC). McKibben type PAMs typically have a threshold pressure of at least lOOkPa and require rigid end fittings which may damage soft tissue and cause local stress concentrations, and thus failure points in the actuator. The actuator design I present is a variant on the McKibben PAM with the following key differences: the nylon mesh is embedded in the elastomeric tube, and closure of the end of the tube is achieved without rigid ends. The actuators were tested to investigate the effects of mesh geometry and elastomer material on force output, contraction, and rise time. Lower initial braid angles and softer elastomer materials provided the best force, contraction, and rise times; Up to 50N of force, 24% contraction, and response times of 0.05s were achieved at 100kPa. The actuators exhibited low threshold pressures (<5kPa) and high rupture pressures (138kPa - 720kPa) which suggest safe operation for the DCC application. These results demonstrate that the actuators can achieve forces, displacements, and rise times suitable to assist with cardiac function.

Degree

thesis:*
Department dc:contributor.department
Massachusetts Institute of Technology. Department of Mechanical Engineering.
Grantor dc:publisher
Massachusetts Institute of Technology
Year dc:date.issued
2013

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Obiajulu, Steven (Steven C.)
Advisor dc:contributor.advisor
  • Daniel D. Frey.

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.
Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1721.1/83730
OAI identifier oai:identifier
oai:dspace.mit.edu:1721.1/83730

Chain of custody

source
Harvested from
MIT
Base URL
dspace.mit.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
citation