Back to results

Massachusetts Institute of Technology

An autocorrelator-interferometer used to determine the pulse width of a pulsed laser used in two-photon endoscopy

Abstract

dc:description.abstract

An autocorrelator-inferometer was designed to correctly assess the pulse width of pulse laser used in two photon endoscopy. The path length of the light was altered using a retro-reflecting corner cube attached to a 6880 galvanometer optical scanner controlled by a 671 series micro-max controller (both products by Cambridge Technologies Inc.) The scanner was selected due to its ability to traverse very small rotations with a constant angular velocity, thereby reducing any non-linearities (with respect to time) in the autocorrelation. The projected results of this autocorrelator suggest it can be used to analyze electromagnetic waves with pulses on the order of a couple picoseconds, however, due to an imbalance of the scanner's shaft, the device was broken before any tests could be performed. A preliminary analysis suggests that a circular shaft attachment could be used to prevent this problem in the future.

Degree

thesis:*
Department dc:contributor.department
Massachusetts Institute of Technology. Dept. of Mechanical Engineering.
Grantor dc:publisher
Massachusetts Institute of Technology
Year dc:date.issued
2005

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Baksh, Nicholas A. (Nicholas Anwar)
Advisor dc:contributor.advisor
  • Peter So.

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.
Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1721.1/32882
OAI identifier oai:identifier
oai:dspace.mit.edu:1721.1/32882

Chain of custody

source
Harvested from
MIT
Base URL
dspace.mit.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
citation

Baksh, Nicholas A. (Nicholas Anwar). An autocorrelator-interferometer used to determine the pulse width of a pulsed laser used in two-photon endoscopy. Massachusetts Institute of Technology, 2005. http://hdl.handle.net/1721.1/32882