Abstract
dc:description.abstract<p>The goal of this study is to design and manufacture a microfluidic device capable of measuring changes in impedance valuesof microfluidic cell cultures. Tocharacterize this, an interdigitated array of electrodes was patterned over glass, where it was then bonded to a series of fluidic networks created in PDMS via soft lithography. The device measured ethanol impedance initially to show that values remain consistent over time. Impedance values of water and 1% wt. saltwater were compared to show that the device is able to detect changes in impedance, with up to a 60% reduction in electrical impedance in saltwater. Cells were introduced into the device, where changes in impedance were seen across multiple frequencies, indicating that the device is capable of detecting the presence of biologic elements within a system. Cell measurements were performed using NIH-3T3 fibroblasts. </p>
Degree
thesis:*- Name thesis:degree_name
- MS in Biomedical Engineering
- Discipline thesis:degree_discipline
- Biomedical and General Engineering
- Year dc:date.available
- 2018
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Foley, John J
- Contributors dc:contributor
-
- Benjamin Hawkins
Subjects
dc:subject × 9Identifiers
dc:identifier.*- Identifier
- 10.15368/theses.2018.146
- OAI identifier oai:identifier
- oai:digitalcommons.calpoly.edu:theses-3343