Back to results

Virginia Tech

Mechanical Comparison of a Type II External Skeletal Fixator and Locking Compression Plate in a Fracture Gap Model

Abstract

dc:description.abstract

The purpose of this study was to compare the stiffness of a Type II external skeletal fixator (ESF) to a 3.5 mm locking compression plate (LCP) in axial compression, mediolateral, and craniocaudal bending in a fracture gap model. The hypothesis was that the Type II ESF would demonstrate comparable stiffness to the LCP. A bone simulant consisting of short fiber reinforced epoxy cylinders and a 40 mm fracture gap was used. The LCP construct consisted of a 12 hole 3.5 mm plate with three 3.5 mm bicortical locking screws per fragment. The Type II ESF construct consisted of 3 proximal full fixation pins (Centerface®) per fragment in the mediolateral plane, and 2 carbon fiber connecting rods. Five constructs of each were tested in non-destructive mediolateral and craniocaudal bending, and axial compression. Stiffness was determined from the slope of the elastic portion of force-displacement curves. A one-way ANOVA and a Tukey-Kramer multiple comparisons test were performed, with significance defined as p < 0.05. In mediolateral bending, the stiffness of the Type II ESF (mean ± standard deviation; 1584.2 N/mm ± 202.8 N/mm) was significantly greater than that of the LCP (110.0 N/mm ± 13.4 N/mm). In axial compression, the stiffness of the Type II ESF (679.1 N/mm ± 20.1 N/mm) was significantly greater than that of the LCP (221.2 N/mm ± 19.1 N/mm). There was no significant difference between the constructs in craniocaudal bending. This information can aid in decision-making for fracture fixation, although ideal stiffness for healing remains unknown.

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Biomedical and Veterinary Sciences
Department dc:contributor.department
Veterinary Medicine
Grantor dc:publisher
Virginia Tech
Year dc:date.issued
2017

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Muro, Noelle Marie
Chair dc:contributor.committeechair
  • Gilley, Robert S.
Committee members dc:contributor.committeemember
  • Kemper, Andrew R.
  • Tempel, Marian Benitez
  • Barry, Sabrina L.

Subjects

dc:subject × 5

Rights

dc:rights
Statement dc:rights
  • In Copyright

Identifiers

dc:identifier.*
Dc Identifier Other
vt_gsexam:10955
OAI identifier oai:identifier
oai:vtechworks.lib.vt.edu:10919/86276

Chain of custody

source
Harvested from
Virginia Tech
Base URL
vtechworks.lib.vt.edu/oai/request
Last updated
2026-07-24
Source record
OAI-PMH GetRecord
citation

Muro, Noelle Marie. Mechanical Comparison of a Type II External Skeletal Fixator and Locking Compression Plate in a Fracture Gap Model. masters thesis, Virginia Tech, 2017. http://hdl.handle.net/10919/86276