Abstract
dc:description.abstractA single crystal neutron spectrometer using a copper crystal has been designed and constructed for the purpose of studying the low energy (<1 ev) neutron spectrum from the V.P.I. reactor. The basic theory necessary for the design, the details of the spectrometer, calibration data and method of operation are presented in this thesis. A relation between angle of diffraction and neutron energy is obtained from the known lattice spacing of the copper crystal. A measure of the total cross section of cadmium from 0.01 to 1.0 ev has been obtained from transmission data for a thin cadmium sheet. The position of the low energy resonance peak confirms the computed calibration. The resolution of the instrument and the intensity of the diffracted beam as a function of energy were studied in order to ascertain the usable range of the spectrometer.
Degree
thesis:*- Name thesis:degree_name
- Master of Science
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- Physics
- Department dc:contributor.department
- Physics
- Grantor dc:publisher
- Virginia Polytechnic Institute
- Year dc:date.issued
- 1963
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Yancey, Kenneth Eugene
Rights
dc:rights- Statement dc:rights
-
- In Copyright
- Licence dc:rights.uri
- Language dc:language.iso
- en_US
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- http://hdl.handle.net/10919/75958
- OAI identifier oai:identifier
- oai:vtechworks.lib.vt.edu:10919/75958