{"id":{"repo_id":"vt","oai_identifier":"oai:vtechworks.lib.vt.edu:10919/75219"},"canonical_url":"https://search.dev.ndltd.org/etd/vt/oai:vtechworks.lib.vt.edu:10919/75219","repository":{"repo_id":"vt","name":"Virginia Tech","base_url":"https://vtechworks.lib.vt.edu/oai/request"},"display":{"title":"Automatic Esophageal Intubation Detection Using Giant Magneto Resistance Sensors","abstract":"This thesis will cover the principle, design, and construction of an automatic esophageal intubation detector. This device uses a giant magneto recitative sensor to and a magnetized stylet to automatically measure the position of an ET tube in a person's throat. This method is less subjective than currently used methods such as end tidal CO2, as it does not rely on user interpretation of data or physiological state of the patient. The device developed during this project was tested on an anatomical mockup, a porcine airway model, and an intubation training dummy. In all three tests, the device performed well, accurately indicating tracheal intubation when the tube was placed in the trachea. Only one instance of a false positive indication of tracheal intubation was recorded and this occurred in an atypical and avoidable situation. As of now, the device functions in non-obese adult male patients, but plans are in place to increase usability for the entire population.","abstract_html":"This thesis will cover the principle, design, and construction of an automatic esophageal intubation detector. This device uses a giant magneto recitative sensor to and a magnetized stylet to automatically measure the position of an ET tube in a person&#x27;s throat. This method is less subjective than currently used methods such as end tidal CO2, as it does not rely on user interpretation of data or physiological state of the patient. The device developed during this project was tested on an anatomical mockup, a porcine airway model, and an intubation training dummy. In all three tests, the device performed well, accurately indicating tracheal intubation when the tube was placed in the trachea. Only one instance of a false positive indication of tracheal intubation was recorded and this occurred in an atypical and avoidable situation. As of now, the device functions in non-obese adult male patients, but plans are in place to increase usability for the entire population.","abstract_has_math":false,"creators":["Alson, Bradley Jacob"],"institution":"Virginia Tech","degree_name":"Master of Science","degree_level":"masters","degree_discipline":"Mechanical Engineering","degree_department":"Mechanical Engineering","school":null,"contributors":[],"advisors":[],"committee_chairs":["Wicks, Alfred L."],"committee_members":["Bird, John P.","Muelenaer, Andre A."],"year":2015,"date_issued":"2015-09-09","date_published":"2015-09-09","updated_at":"2026-07-22T22:18:44Z","subjects":["Airway Management","Intubation"],"languages":[],"rights":["In Copyright"],"rights_urls":["http://rightsstatements.org/vocab/InC/1.0/"],"identifier_entries":[{"key":"dc:identifier.other","label":"Dc Identifier Other","values":["vt_gsexam:6088"],"render_values":[{"text":"vt_gsexam:6088","href":null,"code":true}]}]},"links":{"outbound_url":"http://hdl.handle.net/10919/75219","outbound_label":"Handle","outbound_source":"dc:identifier.uri"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.committeechair","label":"Committee Chair","values":["Wicks, Alfred L."]},{"key":"dc:contributor.committeemember","label":"Committee Member","values":["Bird, John P.","Muelenaer, Andre A."]},{"key":"dc:contributor.department","label":"Department","values":["Mechanical Engineering"]},{"key":"dc:creator","label":"Author","values":["Alson, Bradley Jacob"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2017-03-03T07:00:18Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2017-03-03T07:00:18Z"]},{"key":"dc:date.issued","label":"Date","values":["2015-09-09"]},{"key":"dc:publisher","label":"Institution","values":["Virginia Tech"]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Mechanical Engineering"]},{"key":"thesis:degree_level","label":"Degree Level","values":["masters"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Master of Science"]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["Virginia Polytechnic Institute and State University"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Airway Management","Intubation"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:rights","label":"Dc Rights","values":["In Copyright"]},{"key":"dc:rights.uri","label":"Rights URI","values":["http://rightsstatements.org/vocab/InC/1.0/"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.other","label":"Dc Identifier Other","values":["vt_gsexam:6088"]},{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/10919/75219"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["This thesis will cover the principle, design, and construction of an automatic esophageal intubation detector. This device uses a giant magneto recitative sensor to and a magnetized stylet to automatically measure the position of an ET tube in a person's throat. This method is less subjective than currently used methods such as end tidal CO2, as it does not rely on user interpretation of data or physiological state of the patient. The device developed during this project was tested on an anatomical mockup, a porcine airway model, and an intubation training dummy. In all three tests, the device performed well, accurately indicating tracheal intubation when the tube was placed in the trachea. Only one instance of a false positive indication of tracheal intubation was recorded and this occurred in an atypical and avoidable situation. As of now, the device functions in non-obese adult male patients, but plans are in place to increase usability for the entire population."]},{"key":"dc:description.degree","label":"Dc Description Degree","values":["Master of Science"]},{"key":"dc:format.medium","label":"Dc Format Medium","values":["ETD"]},{"key":"dc:title","label":"Title","values":["Automatic Esophageal Intubation Detection Using Giant Magneto Resistance Sensors"]}]}],"canonical_facts":{"dc:contributor.committeechair":["Wicks, Alfred L."],"dc:contributor.committeemember":["Bird, John P.","Muelenaer, Andre A."],"dc:contributor.department":["Mechanical Engineering"],"dc:creator":["Alson, Bradley Jacob"],"dc:date.accessioned":["2017-03-03T07:00:18Z"],"dc:date.available":["2017-03-03T07:00:18Z"],"dc:date.issued":["2015-09-09"],"dc:description.abstract":["This thesis will cover the principle, design, and construction of an automatic esophageal intubation detector. This device uses a giant magneto recitative sensor to and a magnetized stylet to automatically measure the position of an ET tube in a person's throat. This method is less subjective than currently used methods such as end tidal CO2, as it does not rely on user interpretation of data or physiological state of the patient. The device developed during this project was tested on an anatomical mockup, a porcine airway model, and an intubation training dummy. In all three tests, the device performed well, accurately indicating tracheal intubation when the tube was placed in the trachea. Only one instance of a false positive indication of tracheal intubation was recorded and this occurred in an atypical and avoidable situation. As of now, the device functions in non-obese adult male patients, but plans are in place to increase usability for the entire population."],"dc:description.degree":["Master of Science"],"dc:format.medium":["ETD"],"dc:identifier.other":["vt_gsexam:6088"],"dc:identifier.uri":["http://hdl.handle.net/10919/75219"],"dc:publisher":["Virginia Tech"],"dc:rights":["In Copyright"],"dc:rights.uri":["http://rightsstatements.org/vocab/InC/1.0/"],"dc:subject":["Airway Management","Intubation"],"dc:title":["Automatic Esophageal Intubation Detection Using Giant Magneto Resistance Sensors"],"dc:type":["Thesis"],"thesis:degree_discipline":["Mechanical Engineering"],"thesis:degree_level":["masters"],"thesis:degree_name":["Master of Science"],"thesis:institution_name":["Virginia Polytechnic Institute and State University"]},"updated_at":"2026-07-22T22:18:44Z"}