Back to results

Virginia Tech

A Study of Measuring Intracranial Pressure Using a Non-Invasive Piezoelectric Sensor

Abstract

dc:description.abstract

The brain, like many parts of the human body, can experience swelling, also known as cerebral edema. Cerebral edema may occur because of an injury, health related issues, tumors, or even high altitudes[1]. When cerebral edema occurs, a rise in intracranial pressure (ICP) becomes prevalent and may cause a serious threat. Without immediate treatment, increased intracranial pressure can prevent blood from flowing to the brain and depriving it of necessary oxygen it needs to function. A normal ICP is usually between 5-15 mmHg (666 Pa - 1333Pa). Any ICP observed to be above 20 mmHg (2666Pa) can be associated with brain ischemia and is usually treated[2, 3]. If prolonged, high intracranial pressures can be fatal. Current methods of measuring increased ICP are invasive and may involve drilling into the skull. Extreme invasive measures are not always suitable for certain situations. This thesis presents a study of a non-invasive sensor using piezoelectric PVDF wire to measure the ICP. The PVDF wire sensor is wrapped around the outer portion of the human head to measure the integrated hoop strain. Using this hoop strain, the pressure is then calculated from a known coupling factor of strain to pressure outputted from finite element modeling simulations. The coupling factor is then incorporated into a final calibration factor to calibrate the piezoelectric PVDF wire sensor from charge (Picocoulomb) to pressure (Pascal). These calibration factors are proven to be primarily dependent on the circumference of the human skull. Furthermore, part of this study analyzed the effectiveness and validity of the sensor due to asymmetries in the human skull. A comparison of analytical analysis results versus computational results using finite element modeling simulations show that the PVDF wire sensor neglects any asymmetries presented within the test subject. The results of this study show that this sensor will output correct ICP measurements of different subjects using appropriate calibration factors and is a viable option for measuring ICP non-invasively.

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Mechanical Engineering
Department dc:contributor.department
Mechanical Engineering
Grantor dc:publisher
Virginia Tech
Year dc:date.issued
2014

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Tran, Prenn Xuan
Chair dc:contributor.committeechair
  • Fuller, Christopher R.
Committee members dc:contributor.committeemember
  • Papenfuss, Cory M.
  • Tarazaga, Pablo Alberto

Subjects

dc:subject × 4

Rights

dc:rights
Statement dc:rights
  • In Copyright
Language dc:language.iso
en

Identifiers

dc:identifier.*
Dc Identifier Other
vt_gsexam:3935
OAI identifier oai:identifier
oai:vtechworks.lib.vt.edu:10919/52573

Chain of custody

source
Harvested from
Virginia Tech
Base URL
vtechworks.lib.vt.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
citation

Tran, Prenn Xuan. A Study of Measuring Intracranial Pressure Using a Non-Invasive Piezoelectric Sensor. masters thesis, Virginia Tech, 2014. http://hdl.handle.net/10919/52573