Characterization of Internal Formaldehyde Production within The Pandora Spectrometer Instrument

Abstract

dc:description.abstract

Formaldehyde (HCHO), plays an important role in atmospheric chemistry and is an indicator of atmospheric oxidation capacity and surface ozone photo chemistry. The Pandora Spectrometer Instruments are deployed within the NASA/ESA sponsored Pandonia Global Network designed for satellite validation of various gases in atmosphere (e.g. ozone, nitrogen dioxide and formaldehyde). In addition, Pandoras are extensively used during national (e.g. DISCOVER-AQ, OWLETS, LISTOS) and international (CINDI, KORUS-AQ) field campaigns organized to better characterise air pollution and its distribution. Recently it was discovered and shown in prior research conducted by (Spinei et al. 2020), that Pandora measurements of atmospheric HCHO are impacted by HCHO produced within the telescope assembly due to temperature dependent off-gassing from the Delrin® plastic components. The purpose of the research covered in this thesis is to provide a methodology to correct total HCHO vertical column densities measured during the past field campaigns. The methodology developed through the course of this thesis is first tested on the Pandora simulated measurements derived from the surface concentration HCHO observations during KORUS-AQ (2016) field campaign. The derived correction using synthetic data shows that the proposed methodology is accurate within 30%. The second part of the thesis characterizes heat transfer processes within the telescope assembly to estimate internal temperature as a function of ambient meteorological conditions. Considering that the Pandora instruments have mostly identical design of their telescope assemblies heat transfer coefficients derived from one pandora are expected to be applicable to all Pandoras. Convective heat transfer coefficients were derived at VT wind tunnel as a function of wind speed and telescope assembly position. Internally generated power was measured for several different instruments and averaged at $2.15 pm 0.38$ W. Total long wave emissivity was calculated at 0.63. Surface absorptivities were estimated from the material properties. Semi-empirically derived model is proposed to estimate the internal temperature based on the heat transfer parameters, ambient temperature, relative humidity, solar flux, wind speed and wind direction. The correlation between the estimated and measured internal temperatures is 0.93 R^2. Finally, the methodology is applied to the actual HCHO data collected during the KORUS-AQ campaign and the results are compared to concurrent in-situ measurements made aboard DC-8 aircraft for eight days in the months of May and June 2016.

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Electrical Engineering
Department dc:contributor.department
Electrical Engineering
Grantor dc:publisher
Virginia Tech
Year dc:date.issued
2021

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Kocur, Nash Brinson
Chair dc:contributor.committeechair
  • Lind, Elena Spinei
Committee members dc:contributor.committeemember
  • Zhu, Yizheng
  • Bailey, Scott M.

Subjects

dc:subject × 3

Rights

dc:rights
Statement dc:rights
  • In Copyright

Identifiers

dc:identifier.*
Dc Identifier Other
vt_gsexam:29087
OAI identifier oai:identifier
oai:vtechworks.lib.vt.edu:10919/101973

Chain of custody

source
Harvested from
Virginia Tech
Base URL
vtechworks.lib.vt.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
related terms
citation

Kocur, Nash Brinson. Characterization of Internal Formaldehyde Production within The Pandora Spectrometer Instrument. masters thesis, Virginia Tech, 2021. http://hdl.handle.net/10919/101973