Abstract
dc:description.abstractA transient technique was validated for making thermal conductivity measurements. The technique incorporated a small, effectively spherical, heat source and temperature sensing probe. The actual thermal conductivity measurements lasted 30 seconds. After approximately 15 minutes of data reduction, a value for thermal conductivity was obtained. The probe yielded local thermal conductivity measurements. Spherical sample volumes less than 8 cm² were required for the materials tested. Thermal conductivity (and moisture) distributions can be measured for relatively dry or wetted samples. The technique employs an encapsulated bead thermistor. A thermistor, more commonly used as a temperature transducer, has the inherent feature of being readily self-heated. A computer-based data acquisition and control system regulates the power supplied to the thermistor such that its self-heated temperature response approximates a step change. Thermal conductivity is deduced from the transient measurement of the power dissipated by the probe as a function of time. The technique was used to measure the thermal conductivity of fifteen liquids and five insulation materials. Two different thermistor types, glass-encapsulated and Teflon-encapsulated, were evaluated. Capabilities and limitations of each probe type and the measurement technique, in general, were observed.
Degree
thesis:*- Name thesis:degree_name
- M.S.
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- Mechanical Engineering
- Department dc:contributor.department
- Mechanical Engineering
- Grantor dc:publisher
- Virginia Polytechnic Institute and State University
- Year dc:date.issued
- 1987
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Dougherty, Brian P.
Rights
dc:rights- Statement dc:rights
-
- In Copyright
- Licence dc:rights.uri
- Language dc:language.iso
- en
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- http://hdl.handle.net/10919/101183
- OAI identifier oai:identifier
- oai:vtechworks.lib.vt.edu:10919/101183