Back to results

Virginia Commonwealth University

Design, Fabrication and Performance Evaluation of an Impedimetric Urea Biosensor System

Abstract

dc:description.abstract

An impedance bioanalyzer system comprising an in-vitro biotransducer, instrumentation and control software for the measurement of urea, potentially in blood dialysate, has been developed. The biotransducer comprises of a microlithographically fabricated interdigitated microsensor electrode (IME) onto which was cast a biorecognition layer conferred with the specificity of the enzyme urease. Urease hydrolysis of urea produces NH4+, HC03- and OH- ions that decrease the device's impedance. The temporal rate of change (kinetic) and the extent of change (equilibrium) of ion concentration were measured as the sensor's response. Five formats: [i) unPEGylated urease-containing poly(hydroxyethylmethacrylate) [p(HEMA)] hydrogel, ii) PEGylated urease-containing p(HEMA) hydrogel, iii) via glutaraldehyde crosslinking in the presence of albumin, iv) the direct covalent immobilization of urease to the IME, and v) solution borne urease]. Michaelis-Menten parameters KM, ZMAX and kcat revealed the following rank: PEGylated urease-Gel >> Free Urease > unPEGylated urease-Gel = BSA in Glutaraldehyde > covalently immobilized urease. The unPEGylated-urease sensor provided a higher enzyrne- substrate binding rate and catalysis rate than PEGylated and thus provided a faster impedimetric response to various molar concentrations of urea. Long-term stability (one month) of the PEGylated-urease hydrogel was favorable. A dedicated three-element array impedimetric instrument, the 3EIC BioAnalyzer was designed and produced. A pair of demodulating logarithmic amplifiers (AD8302) was used to calculate the change in phase and amplitude corresponding to the impedimetric response to a 4.0 kHz, 50 mVPP sine wave from a function generator (MAX038). A graphic user interface (GUI), programmed in LabVIEW 7.0 established instrument control, data acquisition via a USB-48A-30A16 μDAQ and graphical data presentation of temporal impedimetric responses.

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
Thesis
Discipline thesis:degree_discipline
Chemical Engineering
Year dc:date.available
2005

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Gupta, Vandana
Contributors dc:contributor
  • Dr. Anthony Guiseppi

Subjects

dc:subject × 11

Rights

dc:rights
Statement dc:rights
  • © The Author

Identifiers

dc:identifier.*
OAI identifier oai:identifier
oai:scholarscompass.vcu.edu:etd-1889

Chain of custody

source
Harvested from
Virginia Commonwealth University
Base URL
scholarscompass.vcu.edu/do/oai/
Last updated
2026-07-24
Source record
OAI-PMH GetRecord
citation

Gupta, Vandana. Design, Fabrication and Performance Evaluation of an Impedimetric Urea Biosensor System. Thesis thesis, 2005. https://doi.org/10.25772/ZRNZ-FK93