Development and Implementation of A Homogeneous and A Heterogeneous Anthropomorphic End to End Quality Assurance Audit System Phantom For Magnetic Resonance Guided Radiotherapy Modalities Ranging From 0.35 T to 1.50 T

Abstract

dc:description.abstract

<p><strong>Introduction: </strong>Magnetic resonance (MR) guided radiation therapy (MRgRT) is an emerging field that integrates an MR imager with either a linear accelerator or three radioactive cobalt-60 sources. Before institutions participate in multi-institutional NCI-sponsored clinical trials, they are required to perform a credentialing test provided by IROC-Houston. During the credentialing test, end-to-end phantoms are used to evaluate the institution’s ability to perform consistent and accurate radiation treatments. IROC-Houston’s conventional anthropomorphic phantoms are not visible in MR, thus they are insufficient for MRgRT systems. The purpose of this work was to create an anthropomorphic thorax and a head and neck (H&N) phantom for MRgRT systems with magnetic fields ranging from 0.35T to 1.5T.</p> <p><strong>Methods: </strong>Over 80 synthetic materials were examined as potential materials used to construct the MRgRT thorax and H&N phantoms. Materials were characterized by: 1) measuring Hounsfield units, 2) visualizing in MR and CT imagers and 3) evaluating their dosimetric characteristics. Once materials were selected for the MRgRT phantoms, radiochromic film and double-loaded TLDs were then characterized in a 1.5T and a 0.35T MR environment. Reproducibility measurements on double-loaded TLDs were performed by using an acrylic block and irradiating it in 0T/1.5T and 0T/0.35T configurations on the Unity system and the MRIdian Cobalt 60 system, respectively. Geometrical thorax and H&N phantom slabs were designed to mimic similar interface conditions seen in anthropomorphic phantoms, but were simplified to reduce manufacturing time. The geometrical phantoms were designed with a rectangular tumor centrally located around surrounding tissue. These two phantoms were used to characterize radiochromic EBT3 film and TLDs by comparing beam profiles and point dose measurements irradiated with and without magnetic fields, respectively. GEANT4 Monte Carlo simulations validated the detectors in both Unity 0T/1.5T and MRIdian 0T/0.35T configurations. Two MRgRT anthropomorphic (H&N and thorax) phantoms were designed, manufactured and evaluated. A reproducibility and feasibility study was conducted to evaluate the phantom’s performance on MRgRT systems.</p> <p><strong>Results: </strong>This study found four materials which were tissue equivalent and visible on both MR and CT. Additionally, this study showed negligible difference in dose response between TLDs and radiochromic film when irradiated in 0.35T and 1.5T magnetic field environments. Two anthropomorphic phantoms were constructed and evaluated. The anthropomorphic thorax and H&N phantoms passed IROC-Houston’s 7%/5mm and 7%/4mm gamma passing criteria, respectively.</p> <p><strong>Conclusions: </strong>An anthropomorphic thorax and an H&N phantom were tissue equivalent, compatible with MR and CT workflows and could be used as end-to-end QA tools for MRgRT systems with magnetic fields ranging from 0.35T to 1.5T.</p>

Degree

thesis:*
Name thesis:degree_name
Doctor of Philosophy (PhD)
Level thesis:degree_level
Dissertation (PhD)
Year dc:date.available
2018

Author and committee

dc:creator, dc:contributor.*
Authors dc:creator
  • Steinmann, Angela
  • <p>0000-0003-4346-1088</p>
Contributors dc:contributor
  • David Followill, Ph.D.
  • Laurence Court, Ph.D.
  • Clifton Fuller, M.D., Ph.D.

Subjects

dc:subject × 14

Identifiers

dc:identifier.*
OAI identifier oai:identifier
oai:digitalcommons.library.tmc.edu:utgsbs_dissertations-1921

Chain of custody

source
Harvested from
University of Texas Health Science Center at Houston
Base URL
digitalcommons.library.tmc.edu/do/oai/
Last updated
2026-07-24
Source record
OAI-PMH GetRecord
citation

Steinmann, Angela; <p>0000-0003-4346-1088</p>. Development and Implementation of A Homogeneous and A Heterogeneous Anthropomorphic End to End Quality Assurance Audit System Phantom For Magnetic Resonance Guided Radiotherapy Modalities Ranging From 0.35 T to 1.50 T. Dissertation (PhD) thesis, 2018. https://digitalcommons.library.tmc.edu/utgsbs_dissertations/876