{"id":{"repo_id":"uthsc","oai_identifier":"oai:digitalcommons.library.tmc.edu:utgsbs_dissertations-1473"},"canonical_url":"https://search.dev.ndltd.org/etd/uthsc/oai:digitalcommons.library.tmc.edu:utgsbs_dissertations-1473","repository":{"repo_id":"uthsc","name":"University of Texas Health Science Center at Houston","base_url":"https://digitalcommons.library.tmc.edu/do/oai/"},"display":{"title":"Use of Positron Emission to mography For Proton Therapy Verification","abstract":"<p>Positron emission tomography (PET), a tool commonly used for cancer staging and response monitoring, has recently been used for proton therapy verification. By imaging tissue activation following proton treatment, attempts have been made to verify proton <em>dose</em> and <em>range</em>. In this dissertation, two novel approaches were developed and tested for the purpose of help improve the proton <em>dose</em> and <em>range </em>estimation as well as verification.</p> <p>Although there are still some challenges, attempts for proton <em>dose</em> verification using PET has been made by comparing Monte Carlo dose and PET simulations with treatment planned dose and measured PET. In this approach, generic tissue composition information is used which can cause large uncertainties in Monte Carlo PET simulation. To improve these uncertainties, we developed a method in the first part of the research to obtain tissue elemental composition using PET after proton therapy. Proton activation of tissue creates progeny radioisotopes according to constituting tissue elements’ cross sections and proton energies. The time-activity-curves from activated tissues can be separated into constituting progeny radioisotopes using a least squares method and then to constituting elemental compositions using Monte Carlo simulated proton energy and cross-section information. We tested this approach using a phantom consisting of sections composed of different combinations of <sup>12</sup>C and <sup>16</sup>O irradiated using a mono-energetic and a SOBP proton beam. In addition, two patient studies were also evaluated using the same technique immediately after proton treatments. The <sup>12</sup>C and <sup>16</sup>O compositions were estimated within 3.6% accuracy in the phantom studies and within 15.2% for patient studies. The obtained tissue elemental composition can be used to improve proton <em>dose</em> verification using PET.</p> <p>In the second part of this dissertation, we developed a new approach to verify proton range. Conventional proton <em>range</em> verification is performed by correlating the distal end of measured PET signals to the proton range. However, this approach is affected by minimal tissue activation near the end of the proton range, perfusion-driven activity washout, and short half-lives of progeny radioisotopes. This also requires an in-beam, in-room, or on-site PET scanner which can be financially and technically challenging for many proton centers. Our new approach overcomes all of these limitations by using proton activated markers. When implanted near the proton distal end, those markers are strongly activated while tissues are minimally activated. For this work we tested <sup>18</sup>O enriched water, Cu and <sup>68</sup>Zn enriched markers that were embedded in tissue-equivalent materials and imaged using an off-site PET scanner following proton activation. The marker materials provided significantly stronger PET signals near the distal end which can be used to verify proton range. In addition, optimal volumes of those markers were also investigated when imbedded in tissue-equivalent materials while using clinical treatment and imaging scenarios. Our results suggest that marker volumes ranging between 5 and 50 mm<sup>3</sup> are required to provide adequate PET signals. The proposed approach can potentially replace conventional fiducial markers with the added benefit of proton <em>range</em> verification.</p> <p>Two proposed methods performed in this research can be used potentially improve both proton <em>dose</em> verification using PET and proton <em>range</em> verification using an off-site PET.</p>","abstract_html":"&lt;p&gt;Positron emission tomography (PET), a tool commonly used for cancer staging and response monitoring, has recently been used for proton therapy verification. By imaging tissue activation following proton treatment, attempts have been made to verify proton &lt;em&gt;dose&lt;/em&gt; and &lt;em&gt;range&lt;/em&gt;. In this dissertation, two novel approaches were developed and tested for the purpose of help improve the proton &lt;em&gt;dose&lt;/em&gt; and &lt;em&gt;range &lt;/em&gt;estimation as well as verification.&lt;/p&gt; &lt;p&gt;Although there are still some challenges, attempts for proton &lt;em&gt;dose&lt;/em&gt; verification using PET has been made by comparing Monte Carlo dose and PET simulations with treatment planned dose and measured PET. In this approach, generic tissue composition information is used which can cause large uncertainties in Monte Carlo PET simulation. To improve these uncertainties, we developed a method in the first part of the research to obtain tissue elemental composition using PET after proton therapy. Proton activation of tissue creates progeny radioisotopes according to constituting tissue elements’ cross sections and proton energies. The time-activity-curves from activated tissues can be separated into constituting progeny radioisotopes using a least squares method and then to constituting elemental compositions using Monte Carlo simulated proton energy and cross-section information. We tested this approach using a phantom consisting of sections composed of different combinations of &lt;sup&gt;12&lt;/sup&gt;C and &lt;sup&gt;16&lt;/sup&gt;O irradiated using a mono-energetic and a SOBP proton beam. In addition, two patient studies were also evaluated using the same technique immediately after proton treatments. The &lt;sup&gt;12&lt;/sup&gt;C and &lt;sup&gt;16&lt;/sup&gt;O compositions were estimated within 3.6% accuracy in the phantom studies and within 15.2% for patient studies. The obtained tissue elemental composition can be used to improve proton &lt;em&gt;dose&lt;/em&gt; verification using PET.&lt;/p&gt; &lt;p&gt;In the second part of this dissertation, we developed a new approach to verify proton range. Conventional proton &lt;em&gt;range&lt;/em&gt; verification is performed by correlating the distal end of measured PET signals to the proton range. However, this approach is affected by minimal tissue activation near the end of the proton range, perfusion-driven activity washout, and short half-lives of progeny radioisotopes. This also requires an in-beam, in-room, or on-site PET scanner which can be financially and technically challenging for many proton centers. Our new approach overcomes all of these limitations by using proton activated markers. When implanted near the proton distal end, those markers are strongly activated while tissues are minimally activated. For this work we tested &lt;sup&gt;18&lt;/sup&gt;O enriched water, Cu and &lt;sup&gt;68&lt;/sup&gt;Zn enriched markers that were embedded in tissue-equivalent materials and imaged using an off-site PET scanner following proton activation. The marker materials provided significantly stronger PET signals near the distal end which can be used to verify proton range. In addition, optimal volumes of those markers were also investigated when imbedded in tissue-equivalent materials while using clinical treatment and imaging scenarios. Our results suggest that marker volumes ranging between 5 and 50 mm&lt;sup&gt;3&lt;/sup&gt; are required to provide adequate PET signals. The proposed approach can potentially replace conventional fiducial markers with the added benefit of proton &lt;em&gt;range&lt;/em&gt; verification.&lt;/p&gt; &lt;p&gt;Two proposed methods performed in this research can be used potentially improve both proton &lt;em&gt;dose&lt;/em&gt; verification using PET and proton &lt;em&gt;range&lt;/em&gt; verification using an off-site PET.&lt;/p&gt;","abstract_has_math":false,"creators":["Cho, Jongmin"],"institution":null,"degree_name":"Doctor of Philosophy (PhD)","degree_level":"Dissertation (PhD)","degree_discipline":null,"degree_department":null,"school":null,"contributors":["Geoffrey Ibbott","Osama Mawlawi","Michael Gillin"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2014,"date_issued":"2014-05-01T07:00:00Z","date_published":"2014-05-01T07:00:00Z","updated_at":"2026-07-24T05:50:31Z","subjects":["positron emission tomography (PET)","proton therapy verification","proton range verification","elemental composition","radioactive decay","implantable marker","proton activation","18O","63Cu","68Zn","Medicine and Health Sciences","Nuclear","Radiology"],"languages":[],"rights":[],"rights_urls":[],"identifier_entries":[]},"links":{"outbound_url":"https://digitalcommons.library.tmc.edu/utgsbs_dissertations/435","outbound_label":"Repository record","outbound_source":"dc:identifier"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Geoffrey Ibbott","Osama Mawlawi","Michael Gillin"]},{"key":"dc:creator","label":"Author","values":["Cho, Jongmin"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.available","label":"Dc Date Available","values":["2015-04-22T07:00:00Z"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Dissertation (PhD)"]},{"key":"thesis:degree_name","label":"Degree Name","values":["Doctor of Philosophy (PhD)"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["positron emission tomography (PET)","proton therapy verification","proton range verification","elemental composition","radioactive decay","implantable marker","proton activation","18O","63Cu","68Zn","Medicine and Health Sciences","Nuclear","Radiology"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["https://digitalcommons.library.tmc.edu/utgsbs_dissertations/435"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description.abstract","label":"Abstract","values":["<p>Positron emission tomography (PET), a tool commonly used for cancer staging and response monitoring, has recently been used for proton therapy verification. By imaging tissue activation following proton treatment, attempts have been made to verify proton <em>dose</em> and <em>range</em>. In this dissertation, two novel approaches were developed and tested for the purpose of help improve the proton <em>dose</em> and <em>range </em>estimation as well as verification.</p> <p>Although there are still some challenges, attempts for proton <em>dose</em> verification using PET has been made by comparing Monte Carlo dose and PET simulations with treatment planned dose and measured PET. In this approach, generic tissue composition information is used which can cause large uncertainties in Monte Carlo PET simulation. To improve these uncertainties, we developed a method in the first part of the research to obtain tissue elemental composition using PET after proton therapy. Proton activation of tissue creates progeny radioisotopes according to constituting tissue elements’ cross sections and proton energies. The time-activity-curves from activated tissues can be separated into constituting progeny radioisotopes using a least squares method and then to constituting elemental compositions using Monte Carlo simulated proton energy and cross-section information. We tested this approach using a phantom consisting of sections composed of different combinations of <sup>12</sup>C and <sup>16</sup>O irradiated using a mono-energetic and a SOBP proton beam. In addition, two patient studies were also evaluated using the same technique immediately after proton treatments. The <sup>12</sup>C and <sup>16</sup>O compositions were estimated within 3.6% accuracy in the phantom studies and within 15.2% for patient studies. The obtained tissue elemental composition can be used to improve proton <em>dose</em> verification using PET.</p> <p>In the second part of this dissertation, we developed a new approach to verify proton range. Conventional proton <em>range</em> verification is performed by correlating the distal end of measured PET signals to the proton range. However, this approach is affected by minimal tissue activation near the end of the proton range, perfusion-driven activity washout, and short half-lives of progeny radioisotopes. This also requires an in-beam, in-room, or on-site PET scanner which can be financially and technically challenging for many proton centers. Our new approach overcomes all of these limitations by using proton activated markers. When implanted near the proton distal end, those markers are strongly activated while tissues are minimally activated. For this work we tested <sup>18</sup>O enriched water, Cu and <sup>68</sup>Zn enriched markers that were embedded in tissue-equivalent materials and imaged using an off-site PET scanner following proton activation. The marker materials provided significantly stronger PET signals near the distal end which can be used to verify proton range. In addition, optimal volumes of those markers were also investigated when imbedded in tissue-equivalent materials while using clinical treatment and imaging scenarios. Our results suggest that marker volumes ranging between 5 and 50 mm<sup>3</sup> are required to provide adequate PET signals. The proposed approach can potentially replace conventional fiducial markers with the added benefit of proton <em>range</em> verification.</p> <p>Two proposed methods performed in this research can be used potentially improve both proton <em>dose</em> verification using PET and proton <em>range</em> verification using an off-site PET.</p>"]},{"key":"dc:title","label":"Title","values":["Use of Positron Emission to mography For Proton Therapy Verification"]}]}],"canonical_facts":{"dc:contributor":["Geoffrey Ibbott","Osama Mawlawi","Michael Gillin"],"dc:creator":["Cho, Jongmin"],"dc:date.available":["2015-04-22T07:00:00Z"],"dc:description.abstract":["<p>Positron emission tomography (PET), a tool commonly used for cancer staging and response monitoring, has recently been used for proton therapy verification. By imaging tissue activation following proton treatment, attempts have been made to verify proton <em>dose</em> and <em>range</em>. In this dissertation, two novel approaches were developed and tested for the purpose of help improve the proton <em>dose</em> and <em>range </em>estimation as well as verification.</p> <p>Although there are still some challenges, attempts for proton <em>dose</em> verification using PET has been made by comparing Monte Carlo dose and PET simulations with treatment planned dose and measured PET. In this approach, generic tissue composition information is used which can cause large uncertainties in Monte Carlo PET simulation. To improve these uncertainties, we developed a method in the first part of the research to obtain tissue elemental composition using PET after proton therapy. Proton activation of tissue creates progeny radioisotopes according to constituting tissue elements’ cross sections and proton energies. The time-activity-curves from activated tissues can be separated into constituting progeny radioisotopes using a least squares method and then to constituting elemental compositions using Monte Carlo simulated proton energy and cross-section information. We tested this approach using a phantom consisting of sections composed of different combinations of <sup>12</sup>C and <sup>16</sup>O irradiated using a mono-energetic and a SOBP proton beam. In addition, two patient studies were also evaluated using the same technique immediately after proton treatments. The <sup>12</sup>C and <sup>16</sup>O compositions were estimated within 3.6% accuracy in the phantom studies and within 15.2% for patient studies. The obtained tissue elemental composition can be used to improve proton <em>dose</em> verification using PET.</p> <p>In the second part of this dissertation, we developed a new approach to verify proton range. Conventional proton <em>range</em> verification is performed by correlating the distal end of measured PET signals to the proton range. However, this approach is affected by minimal tissue activation near the end of the proton range, perfusion-driven activity washout, and short half-lives of progeny radioisotopes. This also requires an in-beam, in-room, or on-site PET scanner which can be financially and technically challenging for many proton centers. Our new approach overcomes all of these limitations by using proton activated markers. When implanted near the proton distal end, those markers are strongly activated while tissues are minimally activated. For this work we tested <sup>18</sup>O enriched water, Cu and <sup>68</sup>Zn enriched markers that were embedded in tissue-equivalent materials and imaged using an off-site PET scanner following proton activation. The marker materials provided significantly stronger PET signals near the distal end which can be used to verify proton range. In addition, optimal volumes of those markers were also investigated when imbedded in tissue-equivalent materials while using clinical treatment and imaging scenarios. Our results suggest that marker volumes ranging between 5 and 50 mm<sup>3</sup> are required to provide adequate PET signals. The proposed approach can potentially replace conventional fiducial markers with the added benefit of proton <em>range</em> verification.</p> <p>Two proposed methods performed in this research can be used potentially improve both proton <em>dose</em> verification using PET and proton <em>range</em> verification using an off-site PET.</p>"],"dc:identifier":["https://digitalcommons.library.tmc.edu/utgsbs_dissertations/435"],"dc:subject":["positron emission tomography (PET)","proton therapy verification","proton range verification","elemental composition","radioactive decay","implantable marker","proton activation","18O","63Cu","68Zn","Medicine and Health Sciences","Nuclear","Radiology"],"dc:title":["Use of Positron Emission to mography For Proton Therapy Verification"],"thesis:degree_level":["Dissertation (PhD)"],"thesis:degree_name":["Doctor of Philosophy (PhD)"]},"updated_at":"2026-07-24T05:50:31Z"}