Comprehensive Calculation-Based Imrt Qa Using R&V Data, Treatment Records, and A Second Treatment Planning System

Abstract

dc:description.abstract

<p><strong>Purpose: </strong>Traditional patient-specific IMRT QA measurements are labor intensive and consume machine time. Calculation-based IMRT QA methods typically are not comprehensive. We have developed a comprehensive calculation-based IMRT QA method to detect uncertainties introduced by the initial dose calculation, the data transfer through the Record-and-Verify (R&V) system, and various aspects of the physical delivery.<strong> </strong></p> <p><strong>Methods: </strong>We recomputed the treatment plans in the patient geometry for 48 cases using data from the R&V, and from the delivery unit to calculate the “as-transferred” and “as-delivered” doses respectively. These data were sent to the original TPS to verify transfer and delivery or to a second TPS to verify the original calculation. For each dataset we examined the dose computed from the R&V record (RV) and from the delivery records (Tx), and the dose computed with a second verification TPS (vTPS). Each verification dose was compared to the clinical dose distribution using 3D gamma analysis and by comparison of mean dose and ROI-specific dose levels to target volumes. Plans were also compared to IMRT QA absolute and relative dose measurements.</p> <p><strong>Results:</strong> The average 3D gamma passing percentages using 3%-3mm, 2%-2mm, and 1%-1mm criteria for the RV plan were 100.0 (σ=0.0), 100.0 (σ=0.0), and 100.0 (σ=0.1); for the Tx plan they were 100.0 (σ=0.0), 100.0 (σ=0.0), and 99.0 (σ=1.4); and for the vTPS plan they were 99.3 (σ=0.6), 97.2 (σ=1.5), and 79.0 (σ=8.6). When comparing target volume doses in the RV, Tx, and vTPS plans to the clinical plans, the average ratios of ROI mean doses were 0.999 (σ=0.001), 1.001 (σ=0.002), and 0.990 (σ=0.009) and ROI-specific dose levels were 0.999 (σ=0.001), 1.001 (σ=0.002), and 0.980 (σ=0.043), respectively. Comparing the clinical, RV, TR, and vTPS calculated doses to the IMRT QA measurements for all 48 patients, the average ratios for absolute doses were 0.999 (σ=0.013), 0.998 (σ=0.013), 0.999 σ=0.015), and 0.990 (σ=0.012), respectively, and the average 2D gamma(5%-3mm) passing percentages for relative doses for 9 patients was were 99.36 (σ=0.68), 99.50 (σ=0.49), 99.13 (σ=0.84), and 98.76 (σ=1.66), respectively.</p> <p><strong>Conclusions:</strong> Together with mechanical and dosimetric QA, our calculation-based IMRT QA method promises to minimize the need for patient-specific QA measurements by identifying outliers in need of further review.</p>

Degree

thesis:*
Name thesis:degree_name
Masters of Science (MS)
Level thesis:degree_level
Thesis (MS)
Year dc:date.available
2012

Author and committee

dc:creator, dc:contributor.*
Authors dc:creator
  • Ohrt, Jared D
  • Balter, Peter
  • Gillin, Michael
  • Laurence, Court
  • Rosen, Issac
  • Johnson, Valen
Contributors dc:contributor
  • Peter Balter, PhD
  • Michael Gillin, PhD
  • Laurence Court, PhD

Subjects

dc:subject × 6

Identifiers

dc:identifier.*
OAI identifier oai:identifier
oai:digitalcommons.library.tmc.edu:utgsbs_dissertations-1308

Chain of custody

source
Harvested from
University of Texas Health Science Center at Houston
Base URL
digitalcommons.library.tmc.edu/do/oai/
Last updated
2026-07-24
Source record
OAI-PMH GetRecord
citation

Ohrt, Jared D; Balter, Peter; Gillin, Michael; Laurence, Court; Rosen, Issac; Johnson, Valen. Comprehensive Calculation-Based Imrt Qa Using R&V Data, Treatment Records, and A Second Treatment Planning System. Thesis (MS) thesis, 2012. https://digitalcommons.library.tmc.edu/utgsbs_dissertations/262