Volumetric Modulated Arc Therapy Evaluation With The Radiological Physics Center Head and Neck Phantom

Abstract

dc:description.abstract

<p>Validation of treatment plan quality and dose calculation accuracy is essential for new radiotherapy techniques, including volumetric modulated arc therapy (VMAT). VMAT delivers intensity modulated radiotherapy treatments while simultaneously rotating the gantry, adding an additional level of complexity to both the dose calculation and delivery of VMAT treatments compared to static gantry IMRT. The purpose of this project was to compare two VMAT systems, Elekta VMAT and Varian RapidArc, to the current standard of care, IMRT, in terms of both treatment plan quality and dosimetric delivery accuracy using the Radiological Physics Center (RPC) head and neck (H&N) phantom. Clinically relevant treatment plans were created for the phantom using typical prescription and dose constraints for Elekta VMAT (planned with Pinnacle3 Smart Arc) and RapidArc and IMRT (both planned with Eclipse). The treatment plans were evaluated to determine if they were clinically comparable using several dosimetric criteria, including ability to meet dose objectives, hot spots, conformity index, and homogeneity index. The planned treatments were delivered to the phantom and absolute doses and relative dose distributions were measured with thermoluminescent dosimeters (TLDs) and radiochromic film, respectively. The measured and calculated doses of each treatment were compared to determine if they were clinically acceptable based upon RPC criteria of ±7% dose difference and 4 mm distance-to-agreement. Gamma analysis was used to assess dosimetric accuracy, as well. All treatment plans were able to meet the dosimetric objectives set by the RPC and had similar hot spots in the normal tissue. The Elekta VMAT plan was more homogenous but less conformal than the RapidArc and IMRT plans. When comparing the measured and calculated doses, all plans met the RPC ±7%/4 mm criteria. The percent of points passing the gamma analysis for each treatment delivery was acceptable. Treatment plan quality of the Elekta VMAT, RapidArc and IMRT treatments were comparable for consistent dose prescriptions and constraints. Additionally, the dosimetric accuracy of the Elekta VMAT and RapidArc treatments was verified to be within acceptable tolerances.</p>

Degree

thesis:*
Name thesis:degree_name
Masters of Science (MS)
Level thesis:degree_level
Thesis (MS)
Year dc:date.available
2010

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Kisling, Kelly
Contributors dc:contributor
  • Rebecca Howell
  • David Followill
  • Geoffrey Ibbott

Subjects

dc:subject × 6

Identifiers

dc:identifier.*
OAI identifier oai:identifier
oai:digitalcommons.library.tmc.edu:utgsbs_dissertations-1093

Chain of custody

source
Harvested from
University of Texas Health Science Center at Houston
Base URL
digitalcommons.library.tmc.edu/do/oai/
Last updated
2026-07-24
Source record
OAI-PMH GetRecord
citation

Kisling, Kelly. Volumetric Modulated Arc Therapy Evaluation With The Radiological Physics Center Head and Neck Phantom. Thesis (MS) thesis, 2010. https://digitalcommons.library.tmc.edu/utgsbs_dissertations/70