Evaluation of The Quantitative Accuracy of A Commercially-Available Positron Emission Mammography Scanner

Abstract

dc:description.abstract

Objective: The PEM Flex Solo II (Naviscan, Inc., San Diego, CA) is currently the only commercially-available positron emission mammography (PEM) scanner. This scanner does not apply corrections for count rate effects, attenuation or scatter during image reconstruction, potentially affecting the quantitative accuracy of images. This work measures the overall quantitative accuracy of the PEM Flex system, and determines the contributions of error due to count rate effects, attenuation and scatter. Materials and Methods: Gelatin phantoms were designed to simulate breasts of different sizes (4 – 12 cm thick) with varying uniform background activity concentration (0.007 – 0.5 μCi/cc), cysts and lesions (2:1, 5:1, 10:1 lesion-to-background ratios). The overall error was calculated from ROI measurements in the phantoms with a clinically relevant background activity concentration (0.065 μCi/cc). The error due to count rate effects was determined by comparing the overall error at multiple background activity concentrations to the error at 0.007 μCi/cc. A point source and cold gelatin phantoms were used to assess the errors due to attenuation and scatter. The maximum pixel values in gelatin and in air were compared to determine the effect of attenuation. Scatter was evaluated by comparing the sum of all pixel values in gelatin and in air. Results: The overall error in the background was found to be negative in phantoms of all thicknesses, with the exception of the 4-cm thick phantoms (0%±7%), and it increased with thickness (-34%±6% for the 12-cm phantoms). All lesions exhibited large negative error (-22% for the 2:1 lesions in the 4-cm phantom) which increased with thickness and with lesion-to-background ratio (-85% for the 10:1 lesions in the 12-cm phantoms). The error due to count rate in phantoms with 0.065 μCi/cc background was negative (-23%±6% for 4-cm thickness) and decreased with thickness (-7%±7% for 12 cm). Attenuation was a substantial source of negative error and increased with thickness (-51%±10% to -77% ±4% in 4 to 12 cm phantoms, respectively). Scatter contributed a relatively constant amount of positive error (+23%±11%) for all thicknesses. Conclusion: Applying corrections for count rate, attenuation and scatter will be essential for the PEM Flex Solo II to be able to produce quantitatively accurate images.

Degree

thesis:*
Name thesis:degree_name
Masters of Science (MS)
Level thesis:degree_level
Thesis (MS)
Year dc:date.available
2010

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Springer, Adam
Contributors dc:contributor
  • Osama Mawlawi, PhD
  • Valen E Johnson, PhD
  • Eric Rohren, MD, PhD

Subjects

dc:subject × 4

Identifiers

dc:identifier.*
OAI identifier oai:identifier
oai:digitalcommons.library.tmc.edu:utgsbs_dissertations-1090

Chain of custody

source
Harvested from
University of Texas Health Science Center at Houston
Base URL
digitalcommons.library.tmc.edu/do/oai/
Last updated
2026-07-24
Source record
OAI-PMH GetRecord
citation

Springer, Adam. Evaluation of The Quantitative Accuracy of A Commercially-Available Positron Emission Mammography Scanner. Thesis (MS) thesis, 2010. https://digitalcommons.library.tmc.edu/utgsbs_dissertations/64