{"id":{"repo_id":"uiuc","oai_identifier":"oai:www.ideals.illinois.edu:2142/110475"},"canonical_url":"https://search.dev.ndltd.org/etd/uiuc/oai:www.ideals.illinois.edu:2142/110475","repository":{"repo_id":"uiuc","name":"University of Illinois - Urbana-Champaign","base_url":"https://www.ideals.illinois.edu/oai-pmh"},"display":{"title":"Real-time beam visualization and therapy planning for focused ultrasound therapies using a diagnostic imaging array","abstract":"Focused ultrasound (FUS) is a non-invasive treatment modality that can produce therapeutic effects in localized regions of tissue through thermal elevations or mechanical stresses caused by an ultrasound wave. Although FUS therapy is effective at treating many different diseases, FUS therapy can be expensive and time-consuming because the treatment must be planned and monitored using guidance technologies such as magnetic resonance imaging (MRI). Pre-therapy planning and real-time monitoring during treatment are crucial to ensure that healthy tissues are not affected by the FUS beam. To provide a cost-effective method for FUS therapy monitoring, we present a novel technique for real-time visualization of an FUS beam using ultrasonic backscatter. A diagnostic imaging array was used to receive backscatter from an FUS beam interacting with tissue and the backscatter was processed to reconstruct the intensity field of the FUS beam. The intensity field reconstruction was then overlaid onto a co-aligned B-mode image captured using the same imaging array to provide anatomical context. To correct for the scattering profile of the medium, the echogenicity of the B-mode image was used to normalize the intensity field reconstruction, allowing for robust beam visualizations even in non-homogeneous media. The beam visualization technique was demonstrated at a frame rate of 25-30 frames per second in a tissue-mimicking phantom and in a rat tumor in vivo while a mock FUS therapy was administered. To facilitate quick and cost-effective therapies, a system for combined therapy planning, real-time beam visualization, and low intensity FUS treatment using a single imaging array was also designed. The therapy planning and beam monitoring capabilities of this system were demonstrated in a tissue-mimicking phantom and in vivo. Passive cavitation detection (PCD) measurements found that an imaging array can induce stable cavitation in microbubbles and could therefore be used to administer some low intensity FUS treatments.","abstract_html":"Focused ultrasound (FUS) is a non-invasive treatment modality that can produce therapeutic effects in localized regions of tissue through thermal elevations or mechanical stresses caused by an ultrasound wave. Although FUS therapy is effective at treating many different diseases, FUS therapy can be expensive and time-consuming because the treatment must be planned and monitored using guidance technologies such as magnetic resonance imaging (MRI). Pre-therapy planning and real-time monitoring during treatment are crucial to ensure that healthy tissues are not affected by the FUS beam. To provide a cost-effective method for FUS therapy monitoring, we present a novel technique for real-time visualization of an FUS beam using ultrasonic backscatter. A diagnostic imaging array was used to receive backscatter from an FUS beam interacting with tissue and the backscatter was processed to reconstruct the intensity field of the FUS beam. The intensity field reconstruction was then overlaid onto a co-aligned B-mode image captured using the same imaging array to provide anatomical context. To correct for the scattering profile of the medium, the echogenicity of the B-mode image was used to normalize the intensity field reconstruction, allowing for robust beam visualizations even in non-homogeneous media. The beam visualization technique was demonstrated at a frame rate of 25-30 frames per second in a tissue-mimicking phantom and in a rat tumor in vivo while a mock FUS therapy was administered. To facilitate quick and cost-effective therapies, a system for combined therapy planning, real-time beam visualization, and low intensity FUS treatment using a single imaging array was also designed. The therapy planning and beam monitoring capabilities of this system were demonstrated in a tissue-mimicking phantom and in vivo. Passive cavitation detection (PCD) measurements found that an imaging array can induce stable cavitation in microbubbles and could therefore be used to administer some low intensity FUS treatments.","abstract_has_math":false,"creators":["Thies, Miles Henry"],"institution":"University of Illinois at Urbana-Champaign","degree_name":"M.S.","degree_level":"Thesis","degree_discipline":"Electrical & Computer Engr","degree_department":null,"school":null,"contributors":["Oelze, Michael L"],"advisors":[],"committee_chairs":[],"committee_members":[],"year":2021,"date_issued":"2021-09-17T01:10:56Z","date_published":"2021-09-17T01:10:56Z","updated_at":"2026-07-22T22:24:50Z","subjects":["Focused ultrasound","Therapy monitoring","Therapy planning"],"languages":["en"],"rights":["Copyright 2021 Miles Thies"],"rights_urls":[],"identifier_entries":[]},"links":{"outbound_url":"http://hdl.handle.net/2142/110475","outbound_label":"Handle","outbound_source":"dc:identifier"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor","label":"Contributor","values":["Oelze, Michael L"]},{"key":"dc:creator","label":"Author","values":["Thies, Miles Henry"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date","label":"Dc Date","values":["2021-09-17T01:10:56Z","2021-04-15","2021-05"]},{"key":"dc:type","label":"Dc Type","values":["text","Thesis"]},{"key":"thesis:degree_discipline","label":"Discipline","values":["Electrical & Computer Engr"]},{"key":"thesis:degree_level","label":"Degree Level","values":["Thesis"]},{"key":"thesis:degree_name","label":"Degree Name","values":["M.S."]},{"key":"thesis:institution_name","label":"Thesis Institution Name","values":["University of Illinois at Urbana-Champaign"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Focused ultrasound","Therapy monitoring","Therapy planning"]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language","label":"Dc Language","values":["en"]},{"key":"dc:rights","label":"Dc Rights","values":["Copyright 2021 Miles Thies"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier","label":"Identifier","values":["http://hdl.handle.net/2142/110475"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Focused ultrasound (FUS) is a non-invasive treatment modality that can produce therapeutic effects in localized regions of tissue through thermal elevations or mechanical stresses caused by an ultrasound wave. Although FUS therapy is effective at treating many different diseases, FUS therapy can be expensive and time-consuming because the treatment must be planned and monitored using guidance technologies such as magnetic resonance imaging (MRI). Pre-therapy planning and real-time monitoring during treatment are crucial to ensure that healthy tissues are not affected by the FUS beam. To provide a cost-effective method for FUS therapy monitoring, we present a novel technique for real-time visualization of an FUS beam using ultrasonic backscatter. A diagnostic imaging array was used to receive backscatter from an FUS beam interacting with tissue and the backscatter was processed to reconstruct the intensity field of the FUS beam. The intensity field reconstruction was then overlaid onto a co-aligned B-mode image captured using the same imaging array to provide anatomical context. To correct for the scattering profile of the medium, the echogenicity of the B-mode image was used to normalize the intensity field reconstruction, allowing for robust beam visualizations even in non-homogeneous media. The beam visualization technique was demonstrated at a frame rate of 25-30 frames per second in a tissue-mimicking phantom and in a rat tumor in vivo while a mock FUS therapy was administered. To facilitate quick and cost-effective therapies, a system for combined therapy planning, real-time beam visualization, and low intensity FUS treatment using a single imaging array was also designed. The therapy planning and beam monitoring capabilities of this system were demonstrated in a tissue-mimicking phantom and in vivo. Passive cavitation detection (PCD) measurements found that an imaging array can induce stable cavitation in microbubbles and could therefore be used to administer some low intensity FUS treatments.","Submission original under an indefinite embargo labeled 'Open Access'. The submission was exported from vireo on 2021-09-16 without embargo terms","The student, Miles Thies, accepted the attached license on 2021-04-14 at 14:42.","The student, Miles Thies, submitted this Thesis for approval on 2021-04-14 at 15:12.","This Thesis was approved for publication on 2021-04-15 at 13:47.","DSpace SAF Submission Ingestion Package generated from Vireo submission #16330 on 2021-09-16 at 16:41:44","Made available in DSpace on 2021-09-17T01:10:56Z (GMT). No. of bitstreams: 43 THIES-THESIS-2021.pdf: 21232044 bytes, checksum: 3776527985b3473ef8679486561dd0b5 (MD5) ICDose.pdf: 42950 bytes, checksum: 14df34a3b438c7ac2aa9b113c72d5582 (MD5) IEEE_ECE.bst: 59476 bytes, checksum: 7668c5e97bcc2d22a9f8d4eab9b269ee (MD5) SCDose.pdf: 45491 bytes, checksum: f511835ff30aaff0f6fed04969c22f21 (MD5) abs.tex: 2052 bytes, checksum: 7b2d2760e39e0745fbb7a6cc4caf2eee (MD5) ack.tex: 673 bytes, checksum: 812ab02c92ef2a6e281552ef1dc82d04 (MD5) animal_a.pdf: 447257 bytes, checksum: af29ce8aebdbf00f1a62556daa0f3a60 (MD5) animal_b.pdf: 494400 bytes, checksum: 7871ce3c80941981d1770018310d2322 (MD5) beam_vis_chapter.tex: 40101 bytes, checksum: fda323afdc84744cbeb851bb1b78a392 (MD5) conclusion.tex: 2754 bytes, checksum: b68b8d38e7e2f81295f74af8c50c8aa9 (MD5) flow_diagram.pdf: 28567 bytes, checksum: f9e0583d8298c5dc0a4c9daf1e58277c (MD5) hydrophone_comparison_a.pdf: 96876 bytes, checksum: 81b9e0cacff80842a7b7ac0303958c1d (MD5) hydrophone_comparison_b.pdf: 528031 bytes, checksum: 497c130d58a523f47ddd0912fed0739a (MD5) hyperechoic_a.pdf: 653555 bytes, checksum: d49e6bb07d4320acf7733efc9476d97c (MD5) hyperechoic_b.pdf: 772429 bytes, checksum: 3fc36580e0354bbc4d5baab44ffb27dd (MD5) hypoechoic_a.pdf: 733099 bytes, checksum: f42124729bee9014ce9af13a1130c168 (MD5) hypoechoic_b.pdf: 748184 bytes, checksum: df3248942dca825ecf5e5033ced6485e (MD5) intro.tex: 10521 bytes, checksum: 0b4c40c631891dad39c1aeaca8e43960 (MD5) main.tex: 5038 bytes, checksum: f08485c3f3201b5d7828425fcc17da6a (MD5) phantom_a.pdf: 454223 bytes, checksum: 703db329c72727273b2dfa60a3402fb9 (MD5) phantom_b.pdf: 461505 bytes, checksum: e75cdbde3016635a58a27ada29cdd030 (MD5) phantom_c.pdf: 758127 bytes, checksum: 378d003b0a573f45e152d7c940de3cdf (MD5) pts_a.pdf: 632774 bytes, checksum: c97a4a0a195e57ad3369eac3c81b4017 (MD5) pts_b.pdf: 739829 bytes, checksum: 0af98c4a9ee2735a68467de3ef10ca76 (MD5) sidelobes.pdf: 30434 bytes, checksum: c9c627b6b2803fb84a81a0d438908574 (MD5) spect_a.pdf: 1400776 bytes, checksum: b63f80e2417b6228c83b3d6199b082bd (MD5) spect_a_control.pdf: 1401163 bytes, checksum: 1dae0d33c80c4b572261dd188ecf3179 (MD5) spect_b.pdf: 1404497 bytes, checksum: 0c8760374035e7499c9eb8d122ec516e (MD5) spect_b_control.pdf: 1402801 bytes, checksum: ec68c581a6b3fb5fbab181843870be96 (MD5) spect_c.pdf: 1398546 bytes, checksum: 99c5a8e2c7bb8e9ff5a4bbf8d8bb6c78 (MD5) spect_c_control.pdf: 1404004 bytes, checksum: 0c07941013aa00fd9b444ac381a0ee95 (MD5) therapyPlan_animal_a.pdf: 265886 bytes, checksum: af57539167d6f5ed7176feb27c9b5dca (MD5) therapyPlan_animal_b.pdf: 273511 bytes, checksum: 65530f4cb417da2ef32d9572655b3ce4 (MD5) therapyPlan_animal_c.pdf: 425635 bytes, checksum: 83374527d30fac6895b37c905740511e (MD5) therapyPlan_flow_diagram.pdf: 30181 bytes, checksum: ab682337505d763dce125e4124febd3e (MD5) therapy_plan_chapter.tex: 26759 bytes, checksum: 0460a5b8e8610f6ab6a840e06f0b7763 (MD5) thesisrefs.bib: 37190 bytes, checksum: ff0679b3b8b8e53dc002d9aeef80227b (MD5) uiucecethesis09.cls: 21498 bytes, checksum: 1d36170e2becb65df2812590bbd4d666 (MD5) wire_target_a.pdf: 1005868 bytes, checksum: b61db49b455f73ab6f68754fd0bd3a0e (MD5) wire_target_b.pdf: 1005166 bytes, checksum: b6426de7ff9af6a7ee0b6c2f598b8916 (MD5) wire_target_c.pdf: 945781 bytes, checksum: e6799ba5be184355ddab8c9554b86225 (MD5) wire_target_d.pdf: 1018289 bytes, checksum: 7824c9fc27c88c4fa8aa082bc70036ad (MD5) LICENSE.txt: 4208 bytes, checksum: 78715927e3e618944445d42315d87c8b (MD5) Previous issue date: 2021-04-15"]},{"key":"dc:format","label":"Dc Format","values":["application/pdf"]},{"key":"dc:title","label":"Title","values":["Real-time beam visualization and therapy planning for focused ultrasound therapies using a diagnostic imaging array"]}]}],"canonical_facts":{"dc:contributor":["Oelze, Michael L"],"dc:creator":["Thies, Miles Henry"],"dc:date":["2021-09-17T01:10:56Z","2021-04-15","2021-05"],"dc:description":["Focused ultrasound (FUS) is a non-invasive treatment modality that can produce therapeutic effects in localized regions of tissue through thermal elevations or mechanical stresses caused by an ultrasound wave. Although FUS therapy is effective at treating many different diseases, FUS therapy can be expensive and time-consuming because the treatment must be planned and monitored using guidance technologies such as magnetic resonance imaging (MRI). Pre-therapy planning and real-time monitoring during treatment are crucial to ensure that healthy tissues are not affected by the FUS beam. To provide a cost-effective method for FUS therapy monitoring, we present a novel technique for real-time visualization of an FUS beam using ultrasonic backscatter. A diagnostic imaging array was used to receive backscatter from an FUS beam interacting with tissue and the backscatter was processed to reconstruct the intensity field of the FUS beam. The intensity field reconstruction was then overlaid onto a co-aligned B-mode image captured using the same imaging array to provide anatomical context. To correct for the scattering profile of the medium, the echogenicity of the B-mode image was used to normalize the intensity field reconstruction, allowing for robust beam visualizations even in non-homogeneous media. The beam visualization technique was demonstrated at a frame rate of 25-30 frames per second in a tissue-mimicking phantom and in a rat tumor in vivo while a mock FUS therapy was administered. To facilitate quick and cost-effective therapies, a system for combined therapy planning, real-time beam visualization, and low intensity FUS treatment using a single imaging array was also designed. The therapy planning and beam monitoring capabilities of this system were demonstrated in a tissue-mimicking phantom and in vivo. Passive cavitation detection (PCD) measurements found that an imaging array can induce stable cavitation in microbubbles and could therefore be used to administer some low intensity FUS treatments.","Submission original under an indefinite embargo labeled 'Open Access'. The submission was exported from vireo on 2021-09-16 without embargo terms","The student, Miles Thies, accepted the attached license on 2021-04-14 at 14:42.","The student, Miles Thies, submitted this Thesis for approval on 2021-04-14 at 15:12.","This Thesis was approved for publication on 2021-04-15 at 13:47.","DSpace SAF Submission Ingestion Package generated from Vireo submission #16330 on 2021-09-16 at 16:41:44","Made available in DSpace on 2021-09-17T01:10:56Z (GMT). No. of bitstreams: 43 THIES-THESIS-2021.pdf: 21232044 bytes, checksum: 3776527985b3473ef8679486561dd0b5 (MD5) ICDose.pdf: 42950 bytes, checksum: 14df34a3b438c7ac2aa9b113c72d5582 (MD5) IEEE_ECE.bst: 59476 bytes, checksum: 7668c5e97bcc2d22a9f8d4eab9b269ee (MD5) SCDose.pdf: 45491 bytes, checksum: f511835ff30aaff0f6fed04969c22f21 (MD5) abs.tex: 2052 bytes, checksum: 7b2d2760e39e0745fbb7a6cc4caf2eee (MD5) ack.tex: 673 bytes, checksum: 812ab02c92ef2a6e281552ef1dc82d04 (MD5) animal_a.pdf: 447257 bytes, checksum: af29ce8aebdbf00f1a62556daa0f3a60 (MD5) animal_b.pdf: 494400 bytes, checksum: 7871ce3c80941981d1770018310d2322 (MD5) beam_vis_chapter.tex: 40101 bytes, checksum: fda323afdc84744cbeb851bb1b78a392 (MD5) conclusion.tex: 2754 bytes, checksum: b68b8d38e7e2f81295f74af8c50c8aa9 (MD5) flow_diagram.pdf: 28567 bytes, checksum: f9e0583d8298c5dc0a4c9daf1e58277c (MD5) hydrophone_comparison_a.pdf: 96876 bytes, checksum: 81b9e0cacff80842a7b7ac0303958c1d (MD5) hydrophone_comparison_b.pdf: 528031 bytes, checksum: 497c130d58a523f47ddd0912fed0739a (MD5) hyperechoic_a.pdf: 653555 bytes, checksum: d49e6bb07d4320acf7733efc9476d97c (MD5) hyperechoic_b.pdf: 772429 bytes, checksum: 3fc36580e0354bbc4d5baab44ffb27dd (MD5) hypoechoic_a.pdf: 733099 bytes, checksum: f42124729bee9014ce9af13a1130c168 (MD5) hypoechoic_b.pdf: 748184 bytes, checksum: df3248942dca825ecf5e5033ced6485e (MD5) intro.tex: 10521 bytes, checksum: 0b4c40c631891dad39c1aeaca8e43960 (MD5) main.tex: 5038 bytes, checksum: f08485c3f3201b5d7828425fcc17da6a (MD5) phantom_a.pdf: 454223 bytes, checksum: 703db329c72727273b2dfa60a3402fb9 (MD5) phantom_b.pdf: 461505 bytes, checksum: e75cdbde3016635a58a27ada29cdd030 (MD5) phantom_c.pdf: 758127 bytes, checksum: 378d003b0a573f45e152d7c940de3cdf (MD5) pts_a.pdf: 632774 bytes, checksum: c97a4a0a195e57ad3369eac3c81b4017 (MD5) pts_b.pdf: 739829 bytes, checksum: 0af98c4a9ee2735a68467de3ef10ca76 (MD5) sidelobes.pdf: 30434 bytes, checksum: c9c627b6b2803fb84a81a0d438908574 (MD5) spect_a.pdf: 1400776 bytes, checksum: b63f80e2417b6228c83b3d6199b082bd (MD5) spect_a_control.pdf: 1401163 bytes, checksum: 1dae0d33c80c4b572261dd188ecf3179 (MD5) spect_b.pdf: 1404497 bytes, checksum: 0c8760374035e7499c9eb8d122ec516e (MD5) spect_b_control.pdf: 1402801 bytes, checksum: ec68c581a6b3fb5fbab181843870be96 (MD5) spect_c.pdf: 1398546 bytes, checksum: 99c5a8e2c7bb8e9ff5a4bbf8d8bb6c78 (MD5) spect_c_control.pdf: 1404004 bytes, checksum: 0c07941013aa00fd9b444ac381a0ee95 (MD5) therapyPlan_animal_a.pdf: 265886 bytes, checksum: af57539167d6f5ed7176feb27c9b5dca (MD5) therapyPlan_animal_b.pdf: 273511 bytes, checksum: 65530f4cb417da2ef32d9572655b3ce4 (MD5) therapyPlan_animal_c.pdf: 425635 bytes, checksum: 83374527d30fac6895b37c905740511e (MD5) therapyPlan_flow_diagram.pdf: 30181 bytes, checksum: ab682337505d763dce125e4124febd3e (MD5) therapy_plan_chapter.tex: 26759 bytes, checksum: 0460a5b8e8610f6ab6a840e06f0b7763 (MD5) thesisrefs.bib: 37190 bytes, checksum: ff0679b3b8b8e53dc002d9aeef80227b (MD5) uiucecethesis09.cls: 21498 bytes, checksum: 1d36170e2becb65df2812590bbd4d666 (MD5) wire_target_a.pdf: 1005868 bytes, checksum: b61db49b455f73ab6f68754fd0bd3a0e (MD5) wire_target_b.pdf: 1005166 bytes, checksum: b6426de7ff9af6a7ee0b6c2f598b8916 (MD5) wire_target_c.pdf: 945781 bytes, checksum: e6799ba5be184355ddab8c9554b86225 (MD5) wire_target_d.pdf: 1018289 bytes, checksum: 7824c9fc27c88c4fa8aa082bc70036ad (MD5) LICENSE.txt: 4208 bytes, checksum: 78715927e3e618944445d42315d87c8b (MD5) Previous issue date: 2021-04-15"],"dc:format":["application/pdf"],"dc:identifier":["http://hdl.handle.net/2142/110475"],"dc:language":["en"],"dc:rights":["Copyright 2021 Miles Thies"],"dc:subject":["Focused ultrasound","Therapy monitoring","Therapy planning"],"dc:title":["Real-time beam visualization and therapy planning for focused ultrasound therapies using a diagnostic imaging array"],"dc:type":["text","Thesis"],"thesis:degree_discipline":["Electrical & Computer Engr"],"thesis:degree_level":["Thesis"],"thesis:degree_name":["M.S."],"thesis:institution_name":["University of Illinois at Urbana-Champaign"]},"updated_at":"2026-07-22T22:24:50Z"}