The development of a remotely expandable anastomosis device for use in paediatric cardiothoracic surgery

Abstract

dc:description.abstract

Congenital Heart Disease (CHD) is one of the main causes of infant mortality in the UK and United States. In 2011, 47% of infants born with CHD in the UK did not survive to their first birthday. Often forms of CHD require the implantation of conduits (tubes) to redirect the blood flow around the reconstructed anatomy in the early days of life. In all cases, these children will require review surgery, often several times to replace the implanted conduit as the child grows. Critically every review procedure carries with it an increased mortality risk.;This project represents the first stage in the development of a conduit/ anastomosis technology, which will grow in response to the growing child, matching haemodynamic demand and eliminating the need for revision surgery.;Two prototypes were designed and developed. Each using an active drive mechanism to expand a stainless steel expansion strip with the slow and controlled movement required for such an application. Each of the designs were silicone coated using custom designed injection moulding techniques ensuring biocompatibility. Power and control systems were designed for each iteration taking into account their expansion methods to achieve the desired expansion rate of 12mm to 20mm in diameter of the implantation period.;Three consecutive implantation studies were carried out throughout the device's development. In the first study, the device was implanted into the right ventricle of a cadaveric pig to determine any potential impediments that would need to be addressed for implantation into live animal models. The second implantation was carried out on 6 piglets, successfully establishing CPB procedures;The third implantation was successfully carried out on 3 piglets, 2 of which survived the duration of the testing, confirming a successful encapsulation method to ensure biocompatibility, however it appeared the device failed to expand due to power line breakdown.

Degree

thesis:*
Name dc:type.qualificationname
phd
Level dc:type.qualificationlevel
doctoral-pg
Grantor dc:publisher.institution
University of Strathclyde
Year dc:date.issued
2017

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Simpson, Craig

Identifiers

dc:identifier.*
Identifier
T14679
Author Identifier
201173675
OAI identifier oai:identifier
oai:strathclyde:7w62f8234

Chain of custody

source
Harvested from
University of Strathclyde
Base URL
stax.strath.ac.uk/catalog/oai
Last updated
2026-07-24
Source record
OAI-PMH GetRecord
related terms
citation

Simpson, Craig. The development of a remotely expandable anastomosis device for use in paediatric cardiothoracic surgery. doctoral-pg thesis, University of Strathclyde, 2017. https://stax.strath.ac.uk/concern/theses/7w62f8234