Abstract

dc:description.abstract

This study investigated an automated classification and diagnosis process for Doppler ultrasound blood flow signals to support clinical decision making. This process is used in this study to diagnose the severity of vascular disease in the lower limb. The study reviewed the existing methods for clinical decision making and proposed a multi-stage process of pre-processing, feature extraction and classification. Each stage was investigated separately and then combined to an automated classification and diagnosis process. Clinical Doppler ultrasound blood flow signals from the lower limb were used to compare existing methods with the novel methods proposed in this study.<br/><br/>The pre-processing stage normally requires the intervention of an operator to select a noise reduction threshold. The novel approach of utilising the Wavelet-based noise reduction method in this study has removed the need for human intervention. Detailed investigations of the feature extraction stage were conducted with the aim to automate and improve the algorithms. Results from these investigations showed that the algorithms proposed in this study, not only automated, but also improved the extraction of the feature vector. Artificial Neural Networks (ANNs) were used to investigate the classification of the Doppler ultrasound blood flow signals into different classes of disease severity.<br/><br/>The results from the investigation into the overall process have proven the full automation of the classification and diagnosis process. This study demonstrated that the introduction of the algorithms proposed in this study significantly improved the diagnosis result over traditional manual methods.<br/>

Degree

thesis:*
Name dc:type.qualificationname
Doctoral Thesis
Level dc:type.qualificationlevel
Student thesis
Year dc:date.issued
2003

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Patz, Ralf

Subjects

dc:subject × 3

Rights

Language dc:language
eng

Identifiers

dc:identifier.*
Identifier
oai:pure.atira.dk:studenttheses/ddd8f346-5f22-4341-a241-3d4dcc0b9f86
OAI identifier oai:identifier
oai:pure.atira.dk:studenttheses/ddd8f346-5f22-4341-a241-3d4dcc0b9f86

Chain of custody

source
Harvested from
University of South Wales
Base URL
pure.southwales.ac.uk/ws/oai
Last updated
2026-07-24
Source record
OAI-PMH GetRecord
citation

Patz, Ralf. Time-frequency Analysis of Doppler Ultrasound Blood Flow Signals: An Investigation into Automated Diagnosis of Vascular Disease. Student thesis thesis, 2003. https://pure.southwales.ac.uk/en/studentTheses/ddd8f346-5f22-4341-a241-3d4dcc0b9f86