Development of a Comprehensive Linac-based Quality Assurance Program for a Retrofitted Micro-MLC SRS System

Abstract

dc:description

In Stereotactic Radiosurgery, one of the most import factors that must be measured is the accuracy of the localization system, whether it be lasers or camera system; and the coincidence of the localized isocenter to the radiation isocenter. According to TG-142 the minimum deviation between the isocenter determined by the localization device and the radiation isocenter of the machine must be no more that 1 mm. In addition, the minimum deviation, also recommended by TG-142, between the radiation isocenter and the mechanical isocenter of the machine must be no more that 1mm.The purpose of this research was to develop a method that both of these parameters could be measured and add these tests to our patient specific QA, monthly QA, and annual QA procedures. A plastic phantom was constructed with holes drilled in each of the sides to meet at a common point in the middle of the phantom. This common intersection was then set as the isocenter for the treatment beams, and the coordinates of the point were sent to the camera system. Measurements were than taken with both the EPID and GafChromic film with the use of rigid tungsten rods in each hole to mark the position of the holes on the film and EPID. The films were then scanned and the field edges and isocenter positions were determined by taking the coordinates of a point that was halfway between the minimum and maximum points in all cases.

Degree

thesis:*
Name thesis:degree_name
Master of Science in Biomedical Sciences (MSBS)
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
College of Medicine
Grantor dc:publisher
University of Toledo Health Science Campus
Year dc:date
2013

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Hancock, George H.
Contributors dc:contributor
  • Pearson, David

Subjects

dc:subject × 2

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center.
Language dc:language
English

Identifiers

dc:identifier.*
OAI identifier oai:identifier
oai:etd.ohiolink.edu:mco1364343813

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-07-24
Source record
OAI-PMH GetRecord
citation

Hancock, George H.. Development of a Comprehensive Linac-based Quality Assurance Program for a Retrofitted Micro-MLC SRS System. masters thesis, University of Toledo Health Science Campus, 2013. http://rave.ohiolink.edu/etdc/view?acc_num=mco1364343813