{"id":{"repo_id":"mit","oai_identifier":"oai:dspace.mit.edu:1721.1/74942"},"canonical_url":"https://search.dev.ndltd.org/etd/mit/oai:dspace.mit.edu:1721.1/74942","repository":{"repo_id":"mit","name":"MIT","base_url":"https://dspace.mit.edu/oai/request"},"display":{"title":"Design and modeling of a PZT thin film based piezoelectric micromachined ultrasonic transducer (PMUT)","abstract":"The design and modelling framework for a piezoelectric micromachined ultrasonic transducer (PMUT) based on the piezoelectric thin film deposition of lead zirconate titanate (PZT) is defined. Through high frequency vibration (1-16MHz) of a thin plate, the PMUT transmits and receives pressure pulses to construct medical ultrasound images as an alternative to bulk piezoelectric transducers currently in use. Existing transducers are difficult to fabricate and lack the small scale necessary for small form factor, high resolution 2D imaging arrays. From acoustic priniciples, the potential PMUT acoustic pressure output is determined and compared to a radiating rigid piston model. Acoustic pressure is shown to scale with the volumetric displacement rate, which is related to the plate deflection. A Green's function approach is then used to explicitly solve for the plate deflection of a bimorph and unimorph PMUT with an arbitrary number of circular or ring electrodes. The resulting solution is much simpler and more flexible than previously published solution techniques enabling the optimization of electrode configuration for large deflection and acoustic pressure. Additionally, the contribution of residual stress is examined; particularly its effects on bandwidth, sensitivity, and resonant frequency and an appropriate electrode coverage of the PMUT plate is suggested. Based on modelling, an initial PMUT design is proposed and is currently being fabricated.","abstract_html":"The design and modelling framework for a piezoelectric micromachined ultrasonic transducer (PMUT) based on the piezoelectric thin film deposition of lead zirconate titanate (PZT) is defined. Through high frequency vibration (1-16MHz) of a thin plate, the PMUT transmits and receives pressure pulses to construct medical ultrasound images as an alternative to bulk piezoelectric transducers currently in use. Existing transducers are difficult to fabricate and lack the small scale necessary for small form factor, high resolution 2D imaging arrays. From acoustic priniciples, the potential PMUT acoustic pressure output is determined and compared to a radiating rigid piston model. Acoustic pressure is shown to scale with the volumetric displacement rate, which is related to the plate deflection. A Green&#x27;s function approach is then used to explicitly solve for the plate deflection of a bimorph and unimorph PMUT with an arbitrary number of circular or ring electrodes. The resulting solution is much simpler and more flexible than previously published solution techniques enabling the optimization of electrode configuration for large deflection and acoustic pressure. Additionally, the contribution of residual stress is examined; particularly its effects on bandwidth, sensitivity, and resonant frequency and an appropriate electrode coverage of the PMUT plate is suggested. Based on modelling, an initial PMUT design is proposed and is currently being fabricated.","abstract_has_math":false,"creators":["Smyth, Katherine Marie"],"institution":"Massachusetts Institute of Technology","degree_name":null,"degree_level":null,"degree_discipline":null,"degree_department":"Massachusetts Institute of Technology. Dept. of Mechanical Engineering.","school":null,"contributors":[],"advisors":["Sang-Gook Kim."],"committee_chairs":[],"committee_members":[],"year":2012,"date_issued":"2012","date_published":"2012","updated_at":"2026-07-22T22:21:34Z","subjects":["Mechanical Engineering."],"languages":["eng"],"rights":["M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission."],"rights_urls":["http://dspace.mit.edu/handle/1721.1/7582"],"identifier_entries":[]},"links":{"outbound_url":"http://hdl.handle.net/1721.1/74942","outbound_label":"Handle","outbound_source":"dc:identifier.uri"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Sang-Gook Kim."]},{"key":"dc:contributor.department","label":"Department","values":["Massachusetts Institute of Technology. Dept. of Mechanical Engineering."]},{"key":"dc:contributor.other","label":"Dc Contributor Other","values":["Massachusetts Institute of Technology. Dept. of Mechanical Engineering."]},{"key":"dc:creator","label":"Author","values":["Smyth, Katherine Marie"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2012-11-19T19:20:28Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2012-11-19T19:20:28Z"]},{"key":"dc:date.issued","label":"Date","values":["2012"]},{"key":"dc:publisher","label":"Institution","values":["Massachusetts Institute of Technology"]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Mechanical Engineering."]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["eng"]},{"key":"dc:rights","label":"Dc Rights","values":["M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission."]},{"key":"dc:rights.uri","label":"Rights URI","values":["http://dspace.mit.edu/handle/1721.1/7582"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/1721.1/74942"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Thesis (S.M.)--Massachusetts Institute of Technology, Dept. of Mechanical Engineering, 2012.","Cataloged from PDF version of thesis.","Includes bibliographical references (p. 149-156)."]},{"key":"dc:description.abstract","label":"Abstract","values":["The design and modelling framework for a piezoelectric micromachined ultrasonic transducer (PMUT) based on the piezoelectric thin film deposition of lead zirconate titanate (PZT) is defined. Through high frequency vibration (1-16MHz) of a thin plate, the PMUT transmits and receives pressure pulses to construct medical ultrasound images as an alternative to bulk piezoelectric transducers currently in use. Existing transducers are difficult to fabricate and lack the small scale necessary for small form factor, high resolution 2D imaging arrays. From acoustic priniciples, the potential PMUT acoustic pressure output is determined and compared to a radiating rigid piston model. Acoustic pressure is shown to scale with the volumetric displacement rate, which is related to the plate deflection. A Green's function approach is then used to explicitly solve for the plate deflection of a bimorph and unimorph PMUT with an arbitrary number of circular or ring electrodes. The resulting solution is much simpler and more flexible than previously published solution techniques enabling the optimization of electrode configuration for large deflection and acoustic pressure. Additionally, the contribution of residual stress is examined; particularly its effects on bandwidth, sensitivity, and resonant frequency and an appropriate electrode coverage of the PMUT plate is suggested. Based on modelling, an initial PMUT design is proposed and is currently being fabricated."]},{"key":"dc:description.degree","label":"Dc Description Degree","values":["S.M."]},{"key":"dc:title","label":"Title","values":["Design and modeling of a PZT thin film based piezoelectric micromachined ultrasonic transducer (PMUT)"]}]}],"canonical_facts":{"dc:contributor.advisor":["Sang-Gook Kim."],"dc:contributor.department":["Massachusetts Institute of Technology. Dept. of Mechanical Engineering."],"dc:contributor.other":["Massachusetts Institute of Technology. Dept. of Mechanical Engineering."],"dc:creator":["Smyth, Katherine Marie"],"dc:date.accessioned":["2012-11-19T19:20:28Z"],"dc:date.available":["2012-11-19T19:20:28Z"],"dc:date.issued":["2012"],"dc:description":["Thesis (S.M.)--Massachusetts Institute of Technology, Dept. of Mechanical Engineering, 2012.","Cataloged from PDF version of thesis.","Includes bibliographical references (p. 149-156)."],"dc:description.abstract":["The design and modelling framework for a piezoelectric micromachined ultrasonic transducer (PMUT) based on the piezoelectric thin film deposition of lead zirconate titanate (PZT) is defined. Through high frequency vibration (1-16MHz) of a thin plate, the PMUT transmits and receives pressure pulses to construct medical ultrasound images as an alternative to bulk piezoelectric transducers currently in use. Existing transducers are difficult to fabricate and lack the small scale necessary for small form factor, high resolution 2D imaging arrays. From acoustic priniciples, the potential PMUT acoustic pressure output is determined and compared to a radiating rigid piston model. Acoustic pressure is shown to scale with the volumetric displacement rate, which is related to the plate deflection. A Green's function approach is then used to explicitly solve for the plate deflection of a bimorph and unimorph PMUT with an arbitrary number of circular or ring electrodes. The resulting solution is much simpler and more flexible than previously published solution techniques enabling the optimization of electrode configuration for large deflection and acoustic pressure. Additionally, the contribution of residual stress is examined; particularly its effects on bandwidth, sensitivity, and resonant frequency and an appropriate electrode coverage of the PMUT plate is suggested. Based on modelling, an initial PMUT design is proposed and is currently being fabricated."],"dc:description.degree":["S.M."],"dc:identifier.uri":["http://hdl.handle.net/1721.1/74942"],"dc:language.iso":["eng"],"dc:publisher":["Massachusetts Institute of Technology"],"dc:rights":["M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission."],"dc:rights.uri":["http://dspace.mit.edu/handle/1721.1/7582"],"dc:subject":["Mechanical Engineering."],"dc:title":["Design and modeling of a PZT thin film based piezoelectric micromachined ultrasonic transducer (PMUT)"],"dc:type":["Thesis"]},"updated_at":"2026-07-22T22:21:34Z"}