Characterization of cardiac lead perforation risk via a dynamic simulated environment

Abstract

dc:description.abstract

Delayed cardiac perforation is a serious medical condition where an implanted cardiac lead migrates through the heart wall, causing life-threatening complications. Where acute perforation occurs during implant, delayed perforation is not detected until after 30 days from implant and is frequently not diagnosed until complications are present. The phenomenon is particularly prevalent in cardiac leads with active helices. Detailed here is a test method that subjects leads to cyclic loading in an environment similar to that of the human heart. A test setup was developed; leads migrate through a degradable cardiac simulant for a fixed cycle count, after which the simulant is evaluated. Initial testing was conducted and improvements made to the device. Nineteen cardiac lead models were tested and ranked in three metrics related to perforation damage. Trends matched available medical knowledge on lead perforation risk. This experiment provides a framework for further investigation into lead perforation.

Degree

thesis:*
Department dc:contributor.department
Massachusetts Institute of Technology. Dept. of Mechanical Engineering.
Grantor dc:publisher
Massachusetts Institute of Technology
Year dc:date.issued
2012

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Rosario, Matthew J
Advisor dc:contributor.advisor
  • Alexander H. Slocum.

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.
Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1721.1/73806
OAI identifier oai:identifier
oai:dspace.mit.edu:1721.1/73806

Chain of custody

source
Harvested from
MIT
Base URL
dspace.mit.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
citation

Rosario, Matthew J. Characterization of cardiac lead perforation risk via a dynamic simulated environment. Massachusetts Institute of Technology, 2012. http://hdl.handle.net/1721.1/73806