{"id":{"repo_id":"mit","oai_identifier":"oai:dspace.mit.edu:1721.1/59954"},"canonical_url":"https://search.dev.ndltd.org/etd/mit/oai:dspace.mit.edu:1721.1/59954","repository":{"repo_id":"mit","name":"MIT","base_url":"https://dspace.mit.edu/oai/request"},"display":{"title":"Design and prototyping of a low-cost portable mechanical ventilator","abstract":"This paper describes the design and prototyping of a low-cost portable mechanical ventilator for use in mass casualty cases and resource-poor environments. The ventilator delivers breaths by compressing a conventional bag-valve mask (BVM) with a pivoting cam arm, eliminating the need for a human operator for the BVM. An initial prototype was built out of acrylic, measuring 11.25 x 6.7 x 8 inches (285 x 170 x 200 mm) and weighing 9 lbs (4.1 kg). It is driven by a stepper motor powered by a 14.8 VDC battery and features an adjustable tidal volume of up to 900 mL, adjustable breaths per minute (bpm) of 5-30, and inhalation to exhalation time ratio (i:e ratio) options of 1:2, 1:3 and 1:4. Tidal volume, breaths per minute and i:e ratio are set via user-friendly knobs, and the settings are displayed on an LCD screen. The prototype also features an assist-control mode and an alarm to indicate over-pressurization of the system. Future iterations of the device will be fully calibrated to medical standards and include all desired ventilator features. Future iterations will be further optimised for low power-consumption and will be designed for manufacture and assembly. With a prototyping cost of only $420, the bulk-manufacturing price for the ventilator is estimated to be less than $100. Through this prototype, the strategy of cam-actuated BVM compression is proven to be a viable option to achieve low-cost, low-power portable ventilator technology that provides essential ventilator features at a fraction of the cost of existing technology. Keywords: Ventilator, Bag Valve Mask (BVM), Low-Cost, Low-Power, Portable and Automatic.","abstract_html":"This paper describes the design and prototyping of a low-cost portable mechanical ventilator for use in mass casualty cases and resource-poor environments. The ventilator delivers breaths by compressing a conventional bag-valve mask (BVM) with a pivoting cam arm, eliminating the need for a human operator for the BVM. An initial prototype was built out of acrylic, measuring 11.25 x 6.7 x 8 inches (285 x 170 x 200 mm) and weighing 9 lbs (4.1 kg). It is driven by a stepper motor powered by a 14.8 VDC battery and features an adjustable tidal volume of up to 900 mL, adjustable breaths per minute (bpm) of 5-30, and inhalation to exhalation time ratio (i:e ratio) options of 1:2, 1:3 and 1:4. Tidal volume, breaths per minute and i:e ratio are set via user-friendly knobs, and the settings are displayed on an LCD screen. The prototype also features an assist-control mode and an alarm to indicate over-pressurization of the system. Future iterations of the device will be fully calibrated to medical standards and include all desired ventilator features. Future iterations will be further optimised for low power-consumption and will be designed for manufacture and assembly. With a prototyping cost of only $420, the bulk-manufacturing price for the ventilator is estimated to be less than $100. Through this prototype, the strategy of cam-actuated BVM compression is proven to be a viable option to achieve low-cost, low-power portable ventilator technology that provides essential ventilator features at a fraction of the cost of existing technology. Keywords: Ventilator, Bag Valve Mask (BVM), Low-Cost, Low-Power, Portable and Automatic.","abstract_has_math":true,"creators":["Powelson, Stephen K. (Stephen Kirby)"],"institution":"Massachusetts Institute of Technology","degree_name":null,"degree_level":null,"degree_discipline":null,"degree_department":"Massachusetts Institute of Technology. Dept. of Mechanical Engineering.","school":null,"contributors":[],"advisors":["Alexander H. Slocum."],"committee_chairs":[],"committee_members":[],"year":2010,"date_issued":"2010","date_published":"2010","updated_at":"2026-07-22T22:22:26Z","subjects":["Mechanical Engineering."],"languages":["eng"],"rights":["M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission."],"rights_urls":["http://dspace.mit.edu/handle/1721.1/7582"],"identifier_entries":[]},"links":{"outbound_url":"http://hdl.handle.net/1721.1/59954","outbound_label":"Handle","outbound_source":"dc:identifier.uri"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["Alexander H. Slocum."]},{"key":"dc:contributor.department","label":"Department","values":["Massachusetts Institute of Technology. Dept. of Mechanical Engineering."]},{"key":"dc:contributor.other","label":"Dc Contributor Other","values":["Massachusetts Institute of Technology. Dept. of Mechanical Engineering."]},{"key":"dc:creator","label":"Author","values":["Powelson, Stephen K. (Stephen Kirby)"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2010-11-08T17:51:13Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2010-11-08T17:51:13Z"]},{"key":"dc:date.issued","label":"Date","values":["2010"]},{"key":"dc:publisher","label":"Institution","values":["Massachusetts Institute of Technology"]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Mechanical Engineering."]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["eng"]},{"key":"dc:rights","label":"Dc Rights","values":["M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission."]},{"key":"dc:rights.uri","label":"Rights URI","values":["http://dspace.mit.edu/handle/1721.1/7582"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/1721.1/59954"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Thesis (S.B.)--Massachusetts Institute of Technology, Dept. of Mechanical Engineering, 2010.","Cataloged from PDF version of thesis.","Includes bibliographical references (p. [10])."]},{"key":"dc:description.abstract","label":"Abstract","values":["This paper describes the design and prototyping of a low-cost portable mechanical ventilator for use in mass casualty cases and resource-poor environments. The ventilator delivers breaths by compressing a conventional bag-valve mask (BVM) with a pivoting cam arm, eliminating the need for a human operator for the BVM. An initial prototype was built out of acrylic, measuring 11.25 x 6.7 x 8 inches (285 x 170 x 200 mm) and weighing 9 lbs (4.1 kg). It is driven by a stepper motor powered by a 14.8 VDC battery and features an adjustable tidal volume of up to 900 mL, adjustable breaths per minute (bpm) of 5-30, and inhalation to exhalation time ratio (i:e ratio) options of 1:2, 1:3 and 1:4. Tidal volume, breaths per minute and i:e ratio are set via user-friendly knobs, and the settings are displayed on an LCD screen. The prototype also features an assist-control mode and an alarm to indicate over-pressurization of the system. Future iterations of the device will be fully calibrated to medical standards and include all desired ventilator features. Future iterations will be further optimised for low power-consumption and will be designed for manufacture and assembly. With a prototyping cost of only $420, the bulk-manufacturing price for the ventilator is estimated to be less than $100. Through this prototype, the strategy of cam-actuated BVM compression is proven to be a viable option to achieve low-cost, low-power portable ventilator technology that provides essential ventilator features at a fraction of the cost of existing technology. Keywords: Ventilator, Bag Valve Mask (BVM), Low-Cost, Low-Power, Portable and Automatic."]},{"key":"dc:description.degree","label":"Dc Description Degree","values":["S.B."]},{"key":"dc:title","label":"Title","values":["Design and prototyping of a low-cost portable mechanical ventilator"]}]}],"canonical_facts":{"dc:contributor.advisor":["Alexander H. Slocum."],"dc:contributor.department":["Massachusetts Institute of Technology. Dept. of Mechanical Engineering."],"dc:contributor.other":["Massachusetts Institute of Technology. Dept. of Mechanical Engineering."],"dc:creator":["Powelson, Stephen K. (Stephen Kirby)"],"dc:date.accessioned":["2010-11-08T17:51:13Z"],"dc:date.available":["2010-11-08T17:51:13Z"],"dc:date.issued":["2010"],"dc:description":["Thesis (S.B.)--Massachusetts Institute of Technology, Dept. of Mechanical Engineering, 2010.","Cataloged from PDF version of thesis.","Includes bibliographical references (p. [10])."],"dc:description.abstract":["This paper describes the design and prototyping of a low-cost portable mechanical ventilator for use in mass casualty cases and resource-poor environments. The ventilator delivers breaths by compressing a conventional bag-valve mask (BVM) with a pivoting cam arm, eliminating the need for a human operator for the BVM. An initial prototype was built out of acrylic, measuring 11.25 x 6.7 x 8 inches (285 x 170 x 200 mm) and weighing 9 lbs (4.1 kg). It is driven by a stepper motor powered by a 14.8 VDC battery and features an adjustable tidal volume of up to 900 mL, adjustable breaths per minute (bpm) of 5-30, and inhalation to exhalation time ratio (i:e ratio) options of 1:2, 1:3 and 1:4. Tidal volume, breaths per minute and i:e ratio are set via user-friendly knobs, and the settings are displayed on an LCD screen. The prototype also features an assist-control mode and an alarm to indicate over-pressurization of the system. Future iterations of the device will be fully calibrated to medical standards and include all desired ventilator features. Future iterations will be further optimised for low power-consumption and will be designed for manufacture and assembly. With a prototyping cost of only $420, the bulk-manufacturing price for the ventilator is estimated to be less than $100. Through this prototype, the strategy of cam-actuated BVM compression is proven to be a viable option to achieve low-cost, low-power portable ventilator technology that provides essential ventilator features at a fraction of the cost of existing technology. Keywords: Ventilator, Bag Valve Mask (BVM), Low-Cost, Low-Power, Portable and Automatic."],"dc:description.degree":["S.B."],"dc:identifier.uri":["http://hdl.handle.net/1721.1/59954"],"dc:language.iso":["eng"],"dc:publisher":["Massachusetts Institute of Technology"],"dc:rights":["M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission."],"dc:rights.uri":["http://dspace.mit.edu/handle/1721.1/7582"],"dc:subject":["Mechanical Engineering."],"dc:title":["Design and prototyping of a low-cost portable mechanical ventilator"],"dc:type":["Thesis"]},"updated_at":"2026-07-22T22:22:26Z"}