Back to results

Massachusetts Institute of Technology

Robotic ultrasound manipulator : calibration of position and orientation measurement system

Abstract

dc:description.abstract

The ultimate goal of this project is to develop a fixture for holding and manipulating an ultrasound wand as adeptly as a trained technician. From the 2D slices recorded by an ultrasound transducer, a three-dimensional image can be reconstructed. To reconstruct the 3D image, position and orientation data are needed. In the interests of space and simplicity, a sensor integrating both a gyroscope and accelerometer was chosen for these measurements. Because of the inherent error induced by integrating acceleration to yield position, it was necessary to calibrate this instrument. A fixture was constructed with preset reference points for the purposes of comparison and error analysis. As expected, the measurements obtained directly from the gyroscope were sufficiently accurate to track orientation. The position values from integration of acceleration exhibited accumulation of error over time. From these data, it was clear that a secondary reference instrument is needed for accurate position measurements.

Degree

thesis:*
Department dc:contributor.department
Massachusetts Institute of Technology. Dept. of Mechanical Engineering.
Grantor dc:publisher
Massachusetts Institute of Technology
Year dc:date.issued
2009

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Syverud, Brian (Brian Chen)
Advisor dc:contributor.advisor
  • Brian W. Anthony.

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.
Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1721.1/54489
OAI identifier oai:identifier
oai:dspace.mit.edu:1721.1/54489

Chain of custody

source
Harvested from
MIT
Base URL
dspace.mit.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
citation

Syverud, Brian (Brian Chen). Robotic ultrasound manipulator : calibration of position and orientation measurement system. Massachusetts Institute of Technology, 2009. http://hdl.handle.net/1721.1/54489