Back to results

Massachusetts Institute of Technology

Monte Carlo simulation of neutron shielding for proton therapy facilities

Abstract

dc:description.abstract

A study was performed to develop a Monte Carlo method of modeling neutron shielding of proton therapy facilities in a complex, realistic environment. The bulk neutron shielding of the Northeast Proton Therapy Center (Massachusetts General Hospital, Boston, MA) was used as the basis of the design work. A geometrical model of the facility was simulated using the LAHET Code System, a set of Monte Carlo codes developed at Los Alamos National Laboratory. Additional software tools for reading and analyzing the simulation data that the model provides have been developed and tested. In order to verify the computer simulations, neutron detection and data acquisition systems have been assembled, modified, and thoroughly tested in order to monitor the neutron dose equivalent during proton beam operation at several locations on a continuous basis. Preliminary tests show that the geometry and physics models proposed in this work are valid.

Degree

thesis:*
Department dc:contributor.department
Massachusetts Institute of Technology. Department of Nuclear Science and Engineering
Grantor dc:publisher
Massachusetts Institute of Technology
Year dc:date.issued
1998

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Folkert, Michael R. (Michael Ryan), 1975-
Advisor dc:contributor.advisor
  • Jacquelyn C. Yanch.

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.
Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1721.1/50492
OAI identifier oai:identifier
oai:dspace.mit.edu:1721.1/50492

Chain of custody

source
Harvested from
MIT
Base URL
dspace.mit.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
related terms
citation

Folkert, Michael R. (Michael Ryan), 1975-. Monte Carlo simulation of neutron shielding for proton therapy facilities. Massachusetts Institute of Technology, 1998. http://hdl.handle.net/1721.1/50492