{"id":{"repo_id":"mit","oai_identifier":"oai:dspace.mit.edu:1721.1/40899"},"canonical_url":"https://search.dev.ndltd.org/etd/mit/oai:dspace.mit.edu:1721.1/40899","repository":{"repo_id":"mit","name":"MIT","base_url":"https://dspace.mit.edu/oai/request"},"display":{"title":"Alcator C-Mod soft X-ray pulse height analysis system","abstract":"A pulse height analysis (PHA) system has been installed on the Alcator C-Mod magnetic confinement fusion experiment. The PHA utilizes a Si(Li) detector to measure soft X-rays in the 1-30 keV range with an energy resolution on the order of 100eV. A FASTComtec MCA-3 multichannel analyzer allows integration times of ims while the PHA supports counting rates of up to 80kHz. The thermal electron temperature is measured from the energy dependence of the intensity of the electron bremsstrahlung spectrum. Measurements of line radiation allows the PHA to identify impurity ions within each plasma shot. The result is a diagnostic that measures thermal electron temperatures while aiding in the indentification of heavy (Z 12) impurities.","abstract_html":"A pulse height analysis (PHA) system has been installed on the Alcator C-Mod magnetic confinement fusion experiment. The PHA utilizes a Si(Li) detector to measure soft X-rays in the 1-30 keV range with an energy resolution on the order of 100eV. A FASTComtec MCA-3 multichannel analyzer allows integration times of ims while the PHA supports counting rates of up to 80kHz. The thermal electron temperature is measured from the energy dependence of the intensity of the electron bremsstrahlung spectrum. Measurements of line radiation allows the PHA to identify impurity ions within each plasma shot. The result is a diagnostic that measures thermal electron temperatures while aiding in the indentification of heavy (Z 12) impurities.","abstract_has_math":false,"creators":["Gamboa, Eliseo (Eliseo J.)"],"institution":"Massachusetts Institute of Technology","degree_name":null,"degree_level":null,"degree_discipline":null,"degree_department":"Massachusetts Institute of Technology. Dept. of Physics.","school":null,"contributors":[],"advisors":["John E. Rice."],"committee_chairs":[],"committee_members":[],"year":2007,"date_issued":"2007","date_published":"2007","updated_at":"2026-07-22T22:22:31Z","subjects":["Physics."],"languages":["eng"],"rights":["M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission."],"rights_urls":["http://dspace.mit.edu/handle/1721.1/7582"],"identifier_entries":[]},"links":{"outbound_url":"http://hdl.handle.net/1721.1/40899","outbound_label":"Handle","outbound_source":"dc:identifier.uri"},"metadata_groups":[{"id":"people","label":"People","entries":[{"key":"dc:contributor.advisor","label":"Advisor","values":["John E. Rice."]},{"key":"dc:contributor.department","label":"Department","values":["Massachusetts Institute of Technology. Dept. of Physics."]},{"key":"dc:contributor.other","label":"Dc Contributor Other","values":["Massachusetts Institute of Technology. Dept. of Physics."]},{"key":"dc:creator","label":"Author","values":["Gamboa, Eliseo (Eliseo J.)"]}]},{"id":"academic_context","label":"Academic Context","entries":[{"key":"dc:date.accessioned","label":"Dc Date Accessioned","values":["2008-03-27T18:21:07Z"]},{"key":"dc:date.available","label":"Dc Date Available","values":["2008-03-27T18:21:07Z"]},{"key":"dc:date.issued","label":"Date","values":["2007"]},{"key":"dc:publisher","label":"Institution","values":["Massachusetts Institute of Technology"]},{"key":"dc:type","label":"Dc Type","values":["Thesis"]}]},{"id":"subjects_keywords","label":"Subjects and Keywords","entries":[{"key":"dc:subject","label":"Dc Subject","values":["Physics."]}]},{"id":"language_rights","label":"Language and Rights","entries":[{"key":"dc:language.iso","label":"Language (ISO)","values":["eng"]},{"key":"dc:rights","label":"Dc Rights","values":["M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission."]},{"key":"dc:rights.uri","label":"Rights URI","values":["http://dspace.mit.edu/handle/1721.1/7582"]}]},{"id":"identifiers","label":"Identifiers","entries":[{"key":"dc:identifier.uri","label":"Identifier URI","values":["http://hdl.handle.net/1721.1/40899"]}]},{"id":"additional","label":"Additional Metadata","entries":[{"key":"dc:description","label":"Description","values":["Thesis (S.B.)--Massachusetts Institute of Technology, Dept. of Physics, 2007.","Includes bibliographical references (leaves 59-60)."]},{"key":"dc:description.abstract","label":"Abstract","values":["A pulse height analysis (PHA) system has been installed on the Alcator C-Mod magnetic confinement fusion experiment. The PHA utilizes a Si(Li) detector to measure soft X-rays in the 1-30 keV range with an energy resolution on the order of 100eV. A FASTComtec MCA-3 multichannel analyzer allows integration times of ims while the PHA supports counting rates of up to 80kHz. The thermal electron temperature is measured from the energy dependence of the intensity of the electron bremsstrahlung spectrum. Measurements of line radiation allows the PHA to identify impurity ions within each plasma shot. The result is a diagnostic that measures thermal electron temperatures while aiding in the indentification of heavy (Z 12) impurities."]},{"key":"dc:description.degree","label":"Dc Description Degree","values":["S.B."]},{"key":"dc:title","label":"Title","values":["Alcator C-Mod soft X-ray pulse height analysis system"]}]}],"canonical_facts":{"dc:contributor.advisor":["John E. Rice."],"dc:contributor.department":["Massachusetts Institute of Technology. Dept. of Physics."],"dc:contributor.other":["Massachusetts Institute of Technology. Dept. of Physics."],"dc:creator":["Gamboa, Eliseo (Eliseo J.)"],"dc:date.accessioned":["2008-03-27T18:21:07Z"],"dc:date.available":["2008-03-27T18:21:07Z"],"dc:date.issued":["2007"],"dc:description":["Thesis (S.B.)--Massachusetts Institute of Technology, Dept. of Physics, 2007.","Includes bibliographical references (leaves 59-60)."],"dc:description.abstract":["A pulse height analysis (PHA) system has been installed on the Alcator C-Mod magnetic confinement fusion experiment. The PHA utilizes a Si(Li) detector to measure soft X-rays in the 1-30 keV range with an energy resolution on the order of 100eV. A FASTComtec MCA-3 multichannel analyzer allows integration times of ims while the PHA supports counting rates of up to 80kHz. The thermal electron temperature is measured from the energy dependence of the intensity of the electron bremsstrahlung spectrum. Measurements of line radiation allows the PHA to identify impurity ions within each plasma shot. The result is a diagnostic that measures thermal electron temperatures while aiding in the indentification of heavy (Z 12) impurities."],"dc:description.degree":["S.B."],"dc:identifier.uri":["http://hdl.handle.net/1721.1/40899"],"dc:language.iso":["eng"],"dc:publisher":["Massachusetts Institute of Technology"],"dc:rights":["M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission."],"dc:rights.uri":["http://dspace.mit.edu/handle/1721.1/7582"],"dc:subject":["Physics."],"dc:title":["Alcator C-Mod soft X-ray pulse height analysis system"],"dc:type":["Thesis"]},"updated_at":"2026-07-22T22:22:31Z"}