Back to results

Massachusetts Institute of Technology

Mitigating Inhomogeneity in High-Field MRI Excitations: Arbitrary Waveform Optimization and Multiphoton Parallel Transmission (MP-pTx)

Abstract

dc:description.abstract

High-field magnetic resonance imaging (MRI) using a standard volume coil results in a spatially varying flip angle across the body, which renders images difficult to clinically interpret. This arises from the complex interactions of electromagnetic fields from current-carrying elements surrounding the imaging region. Parallel transmission (pTx) mitigates this issue by employing multiple high-power, independently controlled transmit elements for more precise excitation control. However, since the wavelength of the applied radio waves is shortened in tissue, the effect becomes highly dependent on the patient’s anatomy. As a result, optimization must be performed on a patient-by-patient basis, and methods that attempt full control of these independent waveforms are too computationally intensive to execute during the limited examination time. Additionally, the high-field excitations create complex electric field distributions that require control and careful monitoring to avoid excessive tissue power deposition (and ultimately heating), quantified as the specific absorption rate (SAR). To address these challenges, we introduce a method for optimizing patient-specific pulses using a global waveform (Ritz) approach, enabling rapid, in-scanner optimization. While pTx effectively addresses flip angle inhomogeneity, it remains costly and introduces challenges in SAR management. We address the SAR management and cost problems of pTx by introducing and characterizing the MP-pTx method, which leverages the multiphoton phenomenon to improve homogeneity using a standard volume coil supplemented with low-frequency (kilohertz) parallel channels. MP-pTx reduces costs and simplifies SAR management by shifting the parallel irradiation to low-cost, lowSAR shim array channels. These channels supplement an off-resonant excitation from a conventional birdcage coil with an oscillating, z-directed field that satisfies the resonance condition for spin state transitions.

Degree

thesis:*
Name thesis:degree_name
Doctoral
Department dc:contributor.department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Grantor dc:publisher
Massachusetts Institute of Technology
Year dc:date.issued
2025

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Drago, John M.
Advisor dc:contributor.advisor
  • Wald, Lawrence L.

Rights

dc:rights
Statement dc:rights
  • In Copyright - Educational Use Permitted
  • Copyright retained by author(s)

Identifiers

dc:identifier.*
Handle dc:identifier.uri
https://hdl.handle.net/1721.1/164039
OAI identifier oai:identifier
oai:dspace.mit.edu:1721.1/164039

Chain of custody

source
Harvested from
MIT
Base URL
dspace.mit.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
related terms
citation

Drago, John M.. Mitigating Inhomogeneity in High-Field MRI Excitations: Arbitrary Waveform Optimization and Multiphoton Parallel Transmission (MP-pTx). Massachusetts Institute of Technology, 2025. https://hdl.handle.net/1721.1/164039