A non-invasive central arterial pressure waveform estimation system using ultrasonography for real-time monitoring

Abstract

dc:description.abstract

This thesis details non-invasive evaluation of a central arterial blood pressure (ABP) waveform using a low-cost ultrasound scanning system. ABP bears significant clinical value in cardiovascular pathophysiology. Non-invasive evaluation of the full ABP waveform has been long desired by medical communities due to its anticipated opportunities to greatly enhance cardiovascular patient care. In addition, central ABP has been focused on because of its close association with the adverse outcomes of cardiovascular events. This work mainly explores monitoring of carotid arterial pulsation and local pulse wave velocity (PWV) by the designed system to estimate the ABP waveform, conducting simultaneous spectral Doppler and M-mode imaging. The system is characterized in electrical and acoustic domains to preserve adequate signal integrity to faithfully extract a spatial mean flow velocity and an arterial distension waveform. The carotid ABP waveform is estimated from the distension waveform and the local PWV with one-time calibration from an arterial-line (A-line) or a volume clamping method. The proof-of-concept study demonstrated that the carotid ABP waveform estimation is feasible. The pulse pressure estimation compared to a sphygmomanometer and a finger ABP waveform differ by 1.49±11.7 and -4.92±12.9 mmHg, respectively. The designed and characterized motion-tolerant ultrasonography extends tolerable lateral offsets up to ±8 mm while limiting error of the flow and distension waveforms within about 5%. The system is also validated under hemodynamic stress of the Valsalva maneuver and in intensive care settings compared to the A-line. This thesis demonstrates the profound potential for a portable low-cost ultrasound system toward non-invasive evaluation of a central ABP waveform in clinically relevant settings.

Degree

thesis:*
Department dc:contributor.department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science.
Grantor dc:publisher
Massachusetts Institute of Technology
Year dc:date.issued
2018

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Seo, Joohyun
Advisor dc:contributor.advisor
  • Hae-Seung Lee and Charles G. Sodini.

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • MIT theses are protected by copyright. They may be viewed, downloaded, or printed from this source but further reproduction or distribution in any format is prohibited without written permission.
Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1721.1/120370
OAI identifier oai:identifier
oai:dspace.mit.edu:1721.1/120370

Chain of custody

source
Harvested from
MIT
Base URL
dspace.mit.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
citation

Seo, Joohyun. A non-invasive central arterial pressure waveform estimation system using ultrasonography for real-time monitoring. Massachusetts Institute of Technology, 2018. http://hdl.handle.net/1721.1/120370