Massachusetts Institute of Technology
Sizing the X-ray spectral resolution limits of the REgolith X-ray Imaging Spectrometer (REXIS) instrument at Asteroid 1999RQ36
Abstract
dc:description.abstractThe REgolith X-ray Imaging Spectrometer (REXIS), a Charge-Coupled Device (CCD)-based coded aperture soft X-ray (0.3-7.5 keV) telescope for remote geochemical X-ray Fluorescence (XRF) spectrometry, will be flying on board the Origins Spectral Interpretations Resource Identification Security Regolith Explorer (OSIRIS-REx) asteroid sample return mission that will be visiting the asteroid 1999 RQ36 and sending a sample back to Earth. REXIS will detect elemental XRF lines and produce a histogram of results as the spacecraft orbits the asteroid as well as produce a global map of elemental abundance ratios. The accuracy requirement for measuring the global ratios of elements and the spectral resolution requirement for discriminating unique XRF lines from each other have been set in place. The correct interpretation of X-ray measurements from the surface of 1999 RQ36 is limited by properties that are intrinsic to the CCD detector, CCID-41, that has been chosen for REXIS. This thesis study outlines student experimentation and results that were conducted on the CCID-41 detector to gauge the intrinsic detector noise as a function of detector temperature. Further, the widening of spectral lines on the resultant histogram was also equated as a function of detector temperature. Members of the REXIS Team built a spectral resolution model to investigate both the widening of spectral lines as a function of detector temperature and the accuracy of the measurement of elemental abundance line ratios as a function of detector temperature. Data from the student laboratory experimentation suggested that the detector temperature remain at or below -75°C to minimize intrinsic noise properties. Data from the computational analyses of the spectral resolution model suggest consistent results that the detector temperature remains at or below -55°C to remain within the established REXIS requirements. The combination of these three results leads to the author's recommendation that a detector temperature requirement be set that the temperature of the CCD detector onboard REXIS shall not exceed -55°C and that a detector temperature goal be set that the temperature of the CCD detector shall not exceed -75°C.
Degree
thesis:*- Department dc:contributor.department
- Massachusetts Institute of Technology. Department of Earth, Atmospheric, and Planetary Sciences.
- Grantor dc:publisher
- Massachusetts Institute of Technology
- Year dc:date.issued
- 2013
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Megyery, Suzanna
- Advisor dc:contributor.advisor
-
- Sara Seager.
Subjects
dc:subject × 1Rights
dc:rights- Statement dc:rights
-
- MIT theses are protected by copyright. They may be viewed, downloaded, or printed from this source but further reproduction or distribution in any format is prohibited without written permission.
- Licence dc:rights.uri
- Language dc:language.iso
- eng
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- http://hdl.handle.net/1721.1/114130
- OAI identifier oai:identifier
- oai:dspace.mit.edu:1721.1/114130