Back to results

Massachusetts Institute of Technology

Electrical system for interfacing with muscle for use in prosthetic devices

Abstract

dc:description.abstract

A shortcoming of current prosthetic technology is the quality of afferent feedback to the user. Modern neural interfaces for powered prosthetic limbs do not provide their users with proprioception, the sense of the position of their limbs in space. This could be achieved by stimulating residual muscles, which contain sensory receptors that are used for proprioception. This scheme requires measurements of muscle length, velocity, and electrical activation in order to provide feedback control for the stimulation. These measurements can also be used to interpret the user's intent for movement of their prosthesis. However, no such hardware exists that is portable, low-power, and that includes functionality for muscle stimulation, length measurement, velocity measurement, and measurement of electrical activity. In this thesis, I designed and built a collection of circuitry for interfacing with muscle through various sensors and stimulation techniques in order to provide this functionality. This system may be used by researchers for taking measurements to deduce muscle state and for closed-loop functional stimulation of muscles. The system includes modules for sonomicrometry, electromyography, and functional electrical stimulation. For each module, the thesis discusses comparable devices, the theory of operation of the device, the design considerations, and and relevant results.

Degree

thesis:*
Department dc:contributor.department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science.
Grantor dc:publisher
Massachusetts Institute of Technology
Year dc:date.issued
2017

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Keyes, Steven Robert
Advisor dc:contributor.advisor
  • Hugh M. Herr.

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • MIT theses are protected by copyright. They may be viewed, downloaded, or printed from this source but further reproduction or distribution in any format is prohibited without written permission.
Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1721.1/113140
OAI identifier oai:identifier
oai:dspace.mit.edu:1721.1/113140

Chain of custody

source
Harvested from
MIT
Base URL
dspace.mit.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
citation

Keyes, Steven Robert. Electrical system for interfacing with muscle for use in prosthetic devices. Massachusetts Institute of Technology, 2017. http://hdl.handle.net/1721.1/113140