Back to results

Department of Medicine

System characterisation of a full body slit scanning radiography machine : theory and experiment

Abstract

dc:description.abstract

This thesis describes a system model of the imaging chain of the Statscan slit scanning full body radiography machine. A Cascaded Linear-Systems model of the detector was developed and the theoretical DQE, MTF and NPS were compared to measure values of the RQA7 beam quality described in IEC 6222-1. The effect on Detector DQE of various system parameters such as coupling efficiency, CCD noise and pixel binning was quantified. Also detector performance for varous thicknesses of Gd2O2S:Tb screens were analysed.

Degree

thesis:*
Grantor dc:publisher.institution
Department of Medicine
Year dc:date.issued
2005

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Scheelke, Martin
Advisor dc:contributor.advisor
  • Douglas, Tania S

Rights

Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/11427/3459
OAI identifier oai:identifier
oai:open.uct.ac.za:11427/3459

Chain of custody

source
Harvested from
University of Cape Town
Base URL
open.uct.ac.za/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
related terms
citation

Scheelke, Martin. System characterisation of a full body slit scanning radiography machine : theory and experiment. Department of Medicine, 2005. http://hdl.handle.net/11427/3459