Back to results

Division of Biomedical Engineering

Radiation dose measurement and prediction for linear slit scanning radiography

Abstract

dc:description.abstract

This study describes dose measurements made for linear slit scanning radiography (LSSR) and a dose prediction model that was developed for LSSR. The measurement and calculation methods used for determining entrance dose and effective dose (E) in conventional X-ray imaging systems were verified for use with LSSR. Entrance dose and E were obtained for LSSR and compared to dose measurements on conventional radiography units. Entrance dose measurements were made using an ionisation chamber and dosemeter; E was calculated from these entrance dose measurements using a Monte Carlo simulator. Comparisons with data from around the world showed that for most examinations the doses obtained for LSSR were considerably lower than those of conventional radiography units for the same image quality. Reasons for the low dose obtained with LSSR include scatter reduction and the beam geometry of LSSR. These results have been published as two papers in international peer reviewed journals. A new method to calculate entrance dose and effective dose for LSSR is described in the second part of this report. This method generates the energy spectrum for a particular set of technique factors, simulates a filter through which the beam is attenuated and then calculates entrance dose directly from this energy spectrum. The energy spectrum is then combined with previously generated organ energy absorption data for a standard sized patient to calculate effective dose to a standard sized patient.Energy imparted for different patient thicknesses can then be used to adjust the effective dose to a patient of any size. This method is performed for a large number of slit beams moving across the body in order to more effectively simulate LSSR. This also allows examinations with technique factors that vary for different parts of the anatomy to be simulated. This method was tested against measured data and Monte Carlo simulations. This model was shown to be accurate, while being specifically suited to LSSR and being considerably faster than Monte Carlo simulations.

Degree

thesis:*
Grantor dc:publisher.institution
Division of Biomedical Engineering
Year dc:date.issued
2008

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Irving, Benjamin
Advisors dc:contributor.advisor
  • Douglas, Tania S
  • Hering, Egbert
  • Maree, Gert

Rights

Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/11427/3251
OAI identifier oai:identifier
oai:open.uct.ac.za:11427/3251

Chain of custody

source
Harvested from
University of Cape Town
Base URL
open.uct.ac.za/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
related terms
citation