Division of Medical Physics
The use of thermoluminescent dosimeters for In-vivo dosimetry in a fast neutron therapy beam
Abstract
dc:description.abstractThermoluminescent detectors (TLD-700) have been investigated for absorbed dose measurements in a p(66)/Be neutron therapy beam at the National Accelerator Centre. Chips were selected based on their reproducibility and chip individual neutron calibration factors were derived. The dose non-linearity was determined in peak 5 and peak 6 and dose non-linearity corrections were performed. The sensitivity of TLD-700 chips with depth and off-axis distance was determined. In-vivo dose measurements were performed on seven patients (9 fields). In the entrance in-vivo dose measurements, a maximal deviation of 3.2 % was detected and a systematic difference of 1.7 % was observed. On the exit side, a maximal deviation of -7.3 % was detected and a systematic difference of -5.1 % was observed. The glow curve peak 6/5 ratio was investigated and found to correlate with the qualitative variations of the average LET in the neutron beam.
Degree
thesis:*- Grantor dc:publisher.institution
- Division of Medical Physics
- Year dc:date.issued
- 2000
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Bhengu, Khumbulani John
- Advisors dc:contributor.advisor
-
- Jones, Dan T L
- Langen, Katja M