Department of Electrical Engineering
Control and monitoring of an intra-aortic balloon pump for cardiac assist
Abstract
dc:description.abstractAn anatomical and physiological background to the heart as a pump, followed by a description of the causes and results of cardiogenic shock is given. A survey of the achievements of other researchers in this field is included. The design of a system for controlling and monitoring the Intra-Aortic Balloon Pump is described. A narrow band-pass filter is used to reduce noise and limit the input to the detector to the spectrum within which the R-wave of an ECG signal lies. A safety circuit to prevent the heart pumping against an inflated balloon is incorporated. History of the 8 secs preceding any instant is available. A description of the medical tests carried out so far is given.
Degree
thesis:*- Grantor dc:publisher.institution
- Department of Electrical Engineering
- Year dc:date.issued
- 1971
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Holmes à Court, Robert Leonard
- Advisor dc:contributor.advisor
-
- Besseling, Johann L N
Rights
- Language dc:language.iso
- eng
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- http://hdl.handle.net/11427/22178
- OAI identifier oai:identifier
- oai:open.uct.ac.za:11427/22178