Back to results

University of Arkansas

Development of an Oxygen Saturation Monitoring System by Embedded Electronics

Abstract

dc:description.abstract

<p>Measuring Oxygenation of blood (SaO2) plays a vital role in patient’s health monitoring. This is often measured by pulse oximeter, which is standard measure during anesthesia, asthma, operative and post-operative recoveries. Despite all, monitoring Oxygen level is necessary for infants with respiratory problems, old people, and pregnant women and in other critical situations. </p> <p>This paper discusses the process of calculating the level of oxygen in blood and heart-rate detection using a non-invasive photo plethysmography also called as pulsoximeter using the MSP430FG437 microcontroller (MCU). The probe uses infrared lights to measure and should be in physical contact with any peripheral points in our body. The percentage of oxygen in the body is worked by measuring the intensity from each frequency of light after it transmits through the body and then calculating the ratio between these two intensities. </p>

Degree

thesis:*
Name thesis:degree_name
Master of Science in Electrical Engineering (MSEE)
Level thesis:degree_level
Thesis
Year dc:date.available
2017

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Venkatesan Gomathy, Manikandan
Advisor dc:contributor.advisor
  • Varadan, Vijay K.
Contributors dc:contributor
  • McCann, Roy A.
  • Ang, Simon S.

Subjects

dc:subject × 4

Identifiers

dc:identifier.*
Repository record dc:identifier
https://scholarworks.uark.edu/etd/1944
OAI identifier oai:identifier
oai:scholarworks.uark.edu:etd-3483

Chain of custody

source
Harvested from
University of Arkansas
Base URL
scholarworks.uark.edu/do/oai/
Last updated
2026-07-24
Source record
OAI-PMH GetRecord
citation

Venkatesan Gomathy, Manikandan. Development of an Oxygen Saturation Monitoring System by Embedded Electronics. Thesis thesis, 2017. https://scholarworks.uark.edu/etd/1944