Virginia Polytechnic Institute and State University
Comparison of high performance liquid chromatographic, gas liquid chromatographic, and Saccharomyces uvarum methods for the determination of B{u2086} compounds
Abstract
dc:description.abstractA high performance liquid chromatographic (HPLC) system consisting of an acetonitrile/phosphate buffer, a Spherisorb ODS column, and UV detector was used to separate the B₆ compounds pyridoxol, pyridoxal, pyridoxamine, and 4'-deoxypyridoxine. A gas chromatography (GC) equipped with a ⁶³Ni electron capture (EC) detector was used to separate the N-methyl-bis-trif luoroacetamide derivatives of the B₆ compounds on a l.54m x 2mm i.d. glass column packed with 10% SP2100 on Supelcoport 90-100 mesh at 125°C and an inlet pressure of 40 psig. Clean and successful separations of all the B₆ forms were obtained by HPLC and GC-EC. Total B₆ values as determined by HPLC and GC-EC for all 3 foods were higher than the corresponding total B₆ values as determined by S. uvarum assay. Several of the B₆ vitamer values for foods obtained by GC-EC agree with corresponding values obtained by HPLC. The HPLC method seemed to be the most satisfactory of the 3 methods for the quantitation of B₆ vitamers in foods and has the following advantages: increased sensitivity, method simplicity, and good precision.
Degree
thesis:*- Name thesis:degree_name
- Ph. D.
- Level thesis:degree_level
- doctoral
- Discipline thesis:degree_discipline
- Human Nutrition and Foods
- Department dc:contributor.department
- Human Nutrition and Foods
- Grantor dc:publisher
- Virginia Polytechnic Institute and State University
- Year dc:date.issued
- 1981
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Lim, Kim L.
Rights
dc:rights- Statement dc:rights
-
- In Copyright
- Licence dc:rights.uri
- Language dc:language.iso
- en_US
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- http://hdl.handle.net/10919/74819
- OAI identifier oai:identifier
- oai:vtechworks.lib.vt.edu:10919/74819