Back to search

Massachusetts Institute of Technology

Drag reducing polymers as simple indicators of hemolytic potential in biomechanical devices

Abstract

dc:description.abstract

An experimental study was carried out to determine if drag reducing polymers can be simple indicators of hemolytic potential in biomechanical devices. Specifically, three different blood pumps, known as a left ventricle assist devices (LVADs) were operated in a test loop using an aqueous solution of polyethylene glycol (PEO, MW = 5000 kDa), a known drag reducing polymer. The pumps were operated under controlled parameters and the change in viscosity (cP) and drag reduction (%DR) for each pump was monitored over the specified time period. The CentriMag® (CM) was used to confirm the drag reducing behavior of PEO, while HeartMate® II (HM II) and HeartMate® III (HM III) were used to determine if there was a correlation between experimental results and actual hemolysis results. Experimental results showed that the mathematical difference between the average final and initial viscosity of HM II was greater than the difference for HM III. HM II had a difference of 0.21 cP and HM III had a difference of 0.16 cP. Hemolysis results using bovine blood showed that HM II had a higher hemolysis rate of 3.80 +/- 1.11 g/day and a higher milligram normalized index of hemolysis of 0.0393 +/- 0.0155. The average hemolysis rate for HM III was 1.38 +/- 0.63 g/day and the milligram normalized index of hemolysis (mg N.I.H.) was 0.571 +/- 0.333. This positive correlation shows that PEO can be a simple indicator of hemolytic potential for biomechanical devices. More data and experimentation is needed to further understand the behavior of PEO and it's ability to indicate hemolytic potential using a wider range of biomechanical devices.

Degree

thesis:*
Department dc:contributor.department
Massachusetts Institute of Technology. Dept. of Mechanical Engineering.
Grantor dc:publisher
Massachusetts Institute of Technology
Year dc:date.issued
2009

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Shieh, Sarah
Advisor dc:contributor.advisor
  • Anette E. Hosoi.

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.
Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1721.1/54495
OAI identifier oai:identifier
oai:dspace.mit.edu:1721.1/54495

Chain of custody

source
Harvested from
MIT
Base URL
dspace.mit.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
citation

Shieh, Sarah. Drag reducing polymers as simple indicators of hemolytic potential in biomechanical devices. Massachusetts Institute of Technology, 2009. http://hdl.handle.net/1721.1/54495