Back to search

Massachusetts Institute of Technology

Off-resonance and detuned surface coils for B₁ inhomogeneity in 7-Tesla MRI

Abstract

dc:description.abstract

A problem with high-field MRI is the lack of B1 homogeneity, particularly signal cancellation in the outer parts of the head. Here we attempt to correct this by adding surface coils. To adjust the mutual coupling, we vary the resonance properties of the added coil. A new agar-based head phantom was built, and two surface coils were built and tuned. The surface coils were placed in various configurations against the phantom to modify the B1 field with their presence, while images were taken using a 16-rung birdcage coil to transmit and receive. Trials were taken with various spacings between the surface coil and the phantom, while the resonance of the surface coil was either shifted in frequency by changing the voltage across a varactor diode, or detuned using a resonant detuning circuit. It was discovered that with a 1 cm spacing and a surface coil tuned just above resonance, SNR near the surface coil could be improved by upwards of 400%, with the trade-off of a reduced signal in other areas on the periphery of the head. Other configurations could achieve better B1 homogeneity at the expense of reduced SNR throughout the head. Future studies will explore the possibility of using more than one surface coil to improve SNR in more places on the periphery of the head.

Degree

thesis:*
Department dc:contributor.department
Massachusetts Institute of Technology. Dept. of Nuclear Science and Engineering.
Grantor dc:publisher
Massachusetts Institute of Technology
Year dc:date.issued
2006

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Zakszewski, Elizabeth K
Advisor dc:contributor.advisor
  • Graham Wiggins.

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.
Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1721.1/41593
OAI identifier oai:identifier
oai:dspace.mit.edu:1721.1/41593

Chain of custody

source
Harvested from
MIT
Base URL
dspace.mit.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
citation

Zakszewski, Elizabeth K. Off-resonance and detuned surface coils for B₁ inhomogeneity in 7-Tesla MRI. Massachusetts Institute of Technology, 2006. http://hdl.handle.net/1721.1/41593