Back to search
Massachusetts Institute of Technology
Lithium-6 filter for a fission converter-based Boron Neutron Capture Therapy irradiation facility beam
Abstract
dc:description.abstract(cont.) A storage system was designed to contain the lithium-6 filter safely when it is not in use. A mixed field dosimetry method was used to measure the photon, thermal neutron and fast neutron dose. The measured advantage depth is 9.3 ± 0.1cm without filter and 9.9 ± 0.1cm with 8mm lithium-6 filter. The result is consistent with the result of Monte Carlo calculation.
Degree
thesis:*- Department dc:contributor.department
- Massachusetts Institute of Technology. Department of Nuclear Engineering
- Grantor dc:publisher
- Massachusetts Institute of Technology
- Year dc:date.issued
- 2005
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Gao, Wei
- Advisor dc:contributor.advisor
-
- Otto K. Harling.
Subjects
dc:subject × 1Rights
dc:rights- Statement dc:rights
-
- M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.
- Licence dc:rights.uri
- Language dc:language.iso
- eng
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- http://hdl.handle.net/1721.1/34653
- OAI identifier oai:identifier
- oai:dspace.mit.edu:1721.1/34653