Massachusetts Institute of Technology

Design of an MRI compatible robot for wrist rehabilitation

Abstract

dc:description.abstract

Magnetic Resonance Imaging (MRI) can be used to study the effects of robotic therapy on stroke patients. This thesis focuses on the design of an MRI compatible wrist robot to be used as a clinical tool for studying the stroke recovery process and the effectiveness of robotic therapy. The robot must comply with strict MRI compatibility requirements, which require careful selection of materials and components. The actuation approach consists of a non- MRI compatible electric motor placed outside the MR room, which is used to drive an MRI compatible hydraulic system. A novel, low friction, MRI compatible hydraulic vane motor was designed for use in the hydraulic system. The robot was designed for use with two interchangeable transmission alternatives: differential gears or friction drive. A stand was designed to allow the wrist robot and patient to fit comfortably within the MRI machine. The MRI compatible wrist robot was assembled and qualitatively tested.

Degree

thesis:*
Department dc:contributor.department
Massachusetts Institute of Technology. Dept. of Mechanical Engineering.
Grantor dc:publisher
Massachusetts Institute of Technology
Year dc:date.issued
2005

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Mendelowitz, Sarah E. (Sarah Elizabeth), 1980-
Advisor dc:contributor.advisor
  • Hermano-Igo Krebs.

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.
Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1721.1/32357
OAI identifier oai:identifier
oai:dspace.mit.edu:1721.1/32357

Chain of custody

source
Harvested from
MIT
Base URL
dspace.mit.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
citation

Mendelowitz, Sarah E. (Sarah Elizabeth), 1980-. Design of an MRI compatible robot for wrist rehabilitation. Massachusetts Institute of Technology, 2005. http://hdl.handle.net/1721.1/32357