Back to search

Massachusetts Institute of Technology

Transcutaneous measurement of glucose using Raman spectroscopy

Abstract

dc:description.abstract

Transcutaneous glucose measurement would provide the ability to obtain frequent measurements without the pain and risk of infection associated with obtaining a blood sample and eliminates the need for reagents. Because of these potential benefits, a large effort is being made within the academic and industrial research communities to develop alternative rapid and easily manageable analysis methods, including transcutaneous methods. Several optical techniques have been proposed and investigated widely to replace the traditional method. Transcutaneous measurement would be of particular benefit to the millions of people with diabetes who should monitor glucose levels multiple times per day and today depend on the finger pricking devices. We have used the advantages provided by NIR Raman spectroscopy, i.e. sharp and distinct spectral features combined with a large probe volume, in order to simultaneously quantify multiple analytes (glucose, urea, cholesterol, triglycerides, total protein, albumin, hemoglobin, hematocrit (hct)) in serum and whole blood samples. Based on the successful measurements, we have moved toward transcutaneous measurements of glucose from 16 healthy human subjects as the first target of blood analytes. RMSEP of 13.17 mg/dL and r² values of 0.79 show promise that the Raman spectroscopy can be developed to achieve clinical accuracy requirements. Furthermore, the fact that glucose features could be recognized in the calibration spectra from measurements of volunteers whose glucose levels changed in a relatively large range, provides evidence that the calibration is based upon glucose.

Degree

thesis:*
Department dc:contributor.department
Massachusetts Institute of Technology. Dept. of Mechanical Engineering.
Grantor dc:publisher
Massachusetts Institute of Technology
Year dc:date.issued
2004

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Oh, Jeankun, 1968-
Advisor dc:contributor.advisor
  • Michael S. Feld.

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.
Language dc:language.iso
en_US

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1721.1/27121
OAI identifier oai:identifier
oai:dspace.mit.edu:1721.1/27121

Chain of custody

source
Harvested from
MIT
Base URL
dspace.mit.edu/oai/request
Last updated
2026-07-22
Source record
OAI-PMH GetRecord
citation

Oh, Jeankun, 1968-. Transcutaneous measurement of glucose using Raman spectroscopy. Massachusetts Institute of Technology, 2004. http://hdl.handle.net/1721.1/27121